滤波反投影重建前预处理过程对低管电压计算机断层成像图像质量的影响。

The effect of a pre-reconstruction process in a filtered back projection reconstruction on an image quality of a low tube voltage computed tomography.

机构信息

Department of Radiological Technology, Yamaguchi University Hospital, Yamaguchi, 755-8505, Japan.

出版信息

Radiol Phys Technol. 2024 Mar;17(1):306-314. doi: 10.1007/s12194-023-00764-9. Epub 2023 Dec 15.

Abstract

This study aims to evaluate the effect of pre-reconstruction process for low tube voltage computed tomography (CT) on image quality of filtered back projection (FBP) reconstruction. Small and large quality assurance water phantoms (19 and 33 cm diameter) were scanned on a third-generation dual-source CT with 70 kVp and 120 kVp at various dose levels. Image quality was assessed in terms of the noise power spectrum (NPS) and task-based transfer function (TTF). NPSs and TTFs in the small phantom were comparable between 70 and 120 kVp protocols. In the large phantom, the curves of the NPS changed and the TTF decreased even at the high-dose levels for 70 kVp protocol compared to 120 kVp protocol. Our results indicated that the pre-reconstruction process is performed in low tube voltage CT for large objects even for the FBP reconstruction and has an effect on the image quality.

摘要

本研究旨在评估低管电压计算机断层扫描(CT)的预重建过程对滤波反投影(FBP)重建图像质量的影响。使用第三代双源 CT 在不同剂量水平下以 70 kVp 和 120 kVp 对小(直径 19 厘米)大和大水模进行扫描。使用噪声功率谱(NPS)和基于任务的传递函数(TTF)评估图像质量。小水模的 70 和 120 kVp 协议之间的 NPS 和 TTF 相当。在大水模中,即使在高剂量水平下,与 120 kVp 协议相比,70 kVp 协议的 NPS 曲线发生变化,TTF 降低。我们的结果表明,即使对于 FBP 重建,大型物体的低管电压 CT 也需要进行预重建过程,并且会对图像质量产生影响。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验