J Hand Surg Eur Vol. 2024 Sep;49(8):987-994. doi: 10.1177/17531934231220637. Epub 2023 Dec 16.
The purpose of this study was to compare computer-aided analysis and different methods of manual measurements in the evaluation of carpal alignment. The radioscaphoid, radiolunate, radiocapitate and radiometacarpal angles were measured on cone-beam computed tomography (CT) scans of 30 healthy wrists by automated software (Disior Ltd.) and by hand surgeons using lateral radiographs reconstructed from the CT data. Hand surgeons were either given ( = 6) or not given ( = 7) prior instructions on how to perform the measurements. Inter- and intra-observer reliability of manual measurements ranged from good to excellent (intra-class correlation coefficients [ICC] 0.77-0.99), being highest in specialists with standardized methods and in reconstructed radiographs with bone overlap digitally removed. Computer-aided software provided excellent intra-observer reliability (ICC 0.94-1.00) consistently and values that were highly comparable (mean difference range 1°-7°) with the manual measurements made in optimal settings. Computer-aided software provides an accurate and repeatable method to measure carpal alignment in CT scans, minimizing observational errors.
本研究旨在比较计算机辅助分析与手动测量方法在腕骨对线评估中的差异。通过自动软件(Disior Ltd.)和手部外科医生使用从 CT 数据重建的侧位 X 线片,对 30 例健康腕关节的锥形束 CT(CT)扫描进行桡腕骨角、桡月骨角、桡头状骨角和桡腕掌角测量。手部外科医生要么( = 6)要么没有( = 7)关于如何进行测量的预先指示。手动测量的组内和组间可靠性范围为良好至极好(组内相关系数 [ICC] 0.77-0.99),在使用标准化方法的专家和数字去除骨重叠的重建 X 线片中最高。计算机辅助软件提供了出色的组内可靠性(ICC 0.94-1.00),并且与在最佳条件下进行的手动测量值高度可比(平均差异范围为 1°-7°)。计算机辅助软件为在 CT 扫描中测量腕骨对线提供了一种准确且可重复的方法,最大限度地减少了观察误差。