To perform an intraindividual comparison of the quality of vision and the effect of decentration between two aspheric intraocular lenses: aspheric balanced curve (ABC) design Vivinex iSert XY1 (Hoya Surgical Optics, Singapore) and anterior aspheric design Tecnis ZCB00 (Abbott Medical Optics, CA).
Tertiary Eye Care Centre.
Prospective, randomized comparative study using a random number table.
Thirty patients were randomized to the implantation of Vivinex iSert XY1 in one eye and Tecnis ZCB00 in the contralateral eye. Then, 12 weeks postoperatively, a laser ray-tracing aberrometer was used to evaluate the visual Strehl ratio, higher-order aberrations (HOA), decentration of IOL from the visual axis and geometric axis, angle alpha, and angle kappa. Contrast sensitivity was measured using the functional visual analyzer.
The visual internal Strehl ratio was higher ( P < 0.05) at all pupil sizes and the spherical aberrations values were lower ( P < 0.05) at larger pupil sizes (5 mm and 6 mm) in the Vivinex group. The mean decentration from the visual axis in the Vivinex group was significantly more than that in the Tecnis group ( P < 0.01). With an increase in angle alpha, there was a greater decline in the visual Strehl ratio for Tecnis; however, there was a statistically significant decline at 3 mm pupil size for Tecnis ZCB00 ( P = 0.028). The contrast sensitivity was similar for both IOLs.
In comparison to an anterior aspheric design IOL, the ABC design IOL yielded better quality of vision, neutralized spherical aberrations to a larger extent, and provided a relatively superior quality of vision with decentration.
对两种非球面人工晶状体(即:非球平衡曲线(ABC)设计 Vivinex iSert XY1(新加坡豪雅 Surgical Optics)和前非球设计 Tecnis ZCB00(加利福尼亚 Abbott Medical Optics))的视觉质量和离焦效果进行个体内比较。
三级眼科保健中心。
使用随机数表的前瞻性、随机对照研究。
将 30 名患者随机分为 Vivinex iSert XY1 单眼植入组和 Tecnis ZCB00 对侧眼植入组。术后 12 周,使用激光射线追踪像差仪评估视觉斯特雷尔比、高阶像差(HOA)、IOL 从视轴和几何轴的离焦、角度α和角度κ。使用功能视觉分析仪测量对比敏感度。
在所有瞳孔大小下,Vivinex 组的视觉内部斯特雷尔比均较高(P < 0.05),在较大瞳孔尺寸(5mm 和 6mm)下,球差值较低(P < 0.05)。Vivinex 组的平均视轴离焦量明显大于 Tecnis 组(P < 0.01)。随着角度α的增加,Tecnis 的视觉斯特雷尔比下降幅度更大;然而,在 Tecnis ZCB00 3mm 瞳孔尺寸时,统计学上有显著下降(P = 0.028)。两种人工晶状体的对比敏感度相似。
与前非球设计人工晶状体相比,ABC 设计人工晶状体产生更好的视觉质量,在更大程度上中和球差,并在离焦时提供相对较好的视觉质量。