Eye (Lond). 2023 Jun;37(8):1640-1645. doi: 10.1038/s41433-022-02211-2. Epub 2022 Aug 24.
BACKGROUND/OBJECTIVES: To determine the influence of decentration and tilt of a pseudophakic aspheric intraocular lens (IOL) on visual acuity (VA) and higher-order aberrations (HOAs), and to analyze the agreement between pupil center/axis and iridocorneal angles center/axis when assessing IOL decentration and tilt.
SUBJECTS/METHODS: A prospective interventional case series study including thirty-three patients undergoing Tecnis ZCB00 (Abbott Medical Optics) implantation. IOL decentration and tilt with respect to two reference systems (pupil and iridocorneal angles centers/axes), in cartesian (X,Y) and polar (radius/tilt, polar angle/azimuth) coordinates, were assessed with optical coherence tomography. VA and internal and ocular HOAs were evaluated. Multiple linear regression models and intraclass correlation coefficient (ICC) were computed.
IOL decentration only showed a significant effect on internal HOAs for [Formula: see text] (R = 0.20, P = 0.04). IOL decentration with respect to the pupil center showed a significant effect on ocular [Formula: see text] (R = 0.18, P = 0.05), [Formula: see text] (R = 0.36, P = 0.001) and [Formula: see text] (R = 0.24, P = 0.02); and with respect to the center of iridocorneal angles, on ocular [Formula: see text] (R = 0.21, P = 0.03), [Formula: see text] (R = 0.32, P = 0.003), primary coma (R = 0.41, P < 0.001), and coma-like (R = 0.40, P = 0.001). Poor agreement between both reference systems was found for IOL decentration measurements (ICC ≤ 0.41), except for the polar angle coordinate (ICC = 0.83). Tilt measurements showed good agreement (ICC ≥ 0.75).
Tecnis ZCB00 decentration and tilt values after uneventful implantation appear not to have influence on VA, and their effect on HOAs are not high enough to clinically affect quality of vision. Pupil and iridocorneal angles used as reference systems may be interchangeable for IOL tilt measurements, but not for decentration.
背景/目的:确定离焦和倾斜的非球面人工晶状体(IOL)对视力(VA)和高阶像差(HOAs)的影响,并分析在评估 IOL 离焦和倾斜时瞳孔中心/轴和虹膜角膜角中心/轴之间的一致性。
受试者/方法:本研究为前瞻性干预性病例系列研究,纳入 33 例接受 Tecnis ZCB00( Abbott Medical Optics)植入的患者。使用光学相干断层扫描评估两个参考系统(瞳孔和虹膜角膜角中心/轴)下 IOL 的离焦和倾斜,以笛卡尔(X,Y)和极坐标(半径/倾斜,极角/方位角)表示。评估 VA 以及内眼和眼内 HOAs。计算多元线性回归模型和组内相关系数(ICC)。
IOL 离焦仅对内眼 HOAs 有显著影响,[公式:见正文](R=0.20,P=0.04)。瞳孔中心的 IOL 离焦对眼内[公式:见正文](R=0.18,P=0.05)、[公式:见正文](R=0.36,P=0.001)和[公式:见正文](R=0.24,P=0.02)有显著影响;而以虹膜角膜角中心为参照,对眼内[公式:见正文](R=0.21,P=0.03)、[公式:见正文](R=0.32,P=0.003)、初级彗差(R=0.41,P<0.001)和彗差样(R=0.40,P=0.001)有显著影响。两种参考系统的 IOL 离焦测量值之间的一致性较差(ICC≤0.41),除极角坐标(ICC=0.83)外。倾斜测量值的一致性较好(ICC≥0.75)。
在无并发症的植入术后,Tecnis ZCB00 的离焦和倾斜值似乎不会影响 VA,对 HOAs 的影响也不足以对视力质量产生临床影响。瞳孔和虹膜角膜角可作为 IOL 倾斜测量的参考系统互换,但不能用于离焦测量。