Huang W Y, Lipsey A I, Cheng M H
Am J Clin Pathol. 1987 Apr;87(4):528-32. doi: 10.1093/ajcp/87.4.528.
The authors have analyzed and compared the cyclosporine concentrations in whole blood specimens from pediatric renal transplant patients using three different methods: high-performance liquid chromatography (HPLC) (5u C18 reverse-phase column), 3H radioimmunoassay (RIA) (Sandoz, E. Hanover, NJ), and 125I RIA (substituted 3H-tracer in Sandoz Kit with 125I tracer made by Immuno Nuclear Corporation, Stillwater, MN). Results obtained by the 125I RIA correlated well with results obtained by the 3H RIA. Both RIA methods had similar correlation with the HPLC method. The 125I RIA method showed higher sensitivity and greater precision than the 3H RIA method. The authors conclude that the 125I RIA method can be used for cyclosporine determination in whole blood specimens. The use of the 125I RIA provides a simple and rapid method with higher counting efficiency and less background quenching than the 3H RIA method, which requires cumbersome liquid scintillation counting procedures.
高效液相色谱法(HPLC)(5u C18反相柱)、3H放射免疫分析法(RIA)(新泽西州伊顿镇山德士公司)以及125I放射免疫分析法(用明尼苏达州斯蒂尔沃特免疫核公司生产的125I示踪剂替代山德士试剂盒中的3H示踪剂)。125I放射免疫分析法得到的结果与3H放射免疫分析法得到的结果相关性良好。两种放射免疫分析法与高效液相色谱法的相关性相似。125I放射免疫分析法比3H放射免疫分析法灵敏度更高、精密度更高。作者得出结论,125I放射免疫分析法可用于全血样本中环孢素的测定。与需要繁琐液体闪烁计数程序的3H放射免疫分析法相比,使用125I放射免疫分析法提供了一种简单快速的方法,具有更高的计数效率和更低的本底猝灭。