Wolf B A, Daft M C, Koenig J W, Flye M W, Turk J W, Scott M G
Department of Pathology, Washington University School of Medicine, St. Louis, MO 63110.
Clin Chem. 1989 Jan;35(1):120-4.
We compared cyclosporine concentrations in whole blood as measured by HPLC and by RIA with a monoclonal antibody specific for cyclosporine with 3H- or 125I-labeled cyclosporine ligand. The 3H-RIA kit slightly underestimated cyclosporine concentrations (greater than 600 micrograms/L) in comparison with HPLC. Over a wide range of concentrations, cyclosporine measured with the 125I-RIA kit correlated well with HPLC (slope = 0.99, n = 301, r = 0.98), observed for samples from recipients of kidney, heart, or liver allografts (respective slopes: 1.01, 0.93, and 1.00). The 125I-RIA standard curve was linear to 1000 micrograms of cyclosporine per liter. Inter- and intra-assay CVs for 125I-RIA measurements of cyclosporine were less than or equal to 7%. Evidently, the 125I-RIA kit involving a monoclonal antibody specific for cyclosporine is equivalent to the HPLC assay and can replace it for therapeutic drug monitoring of cyclosporine therapy.
我们比较了通过高效液相色谱法(HPLC)和放射免疫分析法(RIA)(使用针对环孢素的单克隆抗体与3H或125I标记的环孢素配体)测定的全血中环孢素浓度。与HPLC相比,3H-RIA试剂盒略微低估了环孢素浓度(大于600微克/升)。在很宽的浓度范围内,用125I-RIA试剂盒测定的环孢素与HPLC相关性良好(斜率 = 0.99,n = 301,r = 0.98),这在肾、心或肝同种异体移植受者的样本中观察到(各自斜率:1.01、0.93和1.00)。125I-RIA标准曲线在每升1000微克环孢素范围内呈线性。125I-RIA测定环孢素的批间和批内变异系数小于或等于7%。显然,涉及针对环孢素的单克隆抗体的125I-RIA试剂盒等同于HPLC测定法,并且可以替代它用于环孢素治疗的治疗药物监测。