Blood Press Monit. 2024 Apr 1;29(2):82-88. doi: 10.1097/MBP.0000000000000689. Epub 2024 Jan 15.
The objective of this study was to assess the feasibility of the Arteriograph 24 device to measure 24-hour PWV and central systolic blood pressure (cSBP) in patients with type 2 diabetes (T2DM) and non-diabetic controls and compare daytime and nighttime characteristics in the two groups. Twenty-four-hour PWV and cSBP was measured in 58 patients with T2DM (mean age: 66 ± 9 years, 50% women, mean duration of T2DM: 7.8 ± 1.5 years) and 62 age- and sex-matched controls. Seventy percent of participants (71% T2DM patients and 69% controls) had sufficient readings to generate an acceptable 24-hour report (≥14 day and ≥7 night readings). Lower nocturnal than daytime PWV and cSBP were observed in both groups. Nocturnal PWV and cSBP dipping were attenuated in T2DM patients compared to controls (PWV: -0.3 ± 0.9 vs. -0.7 ± 0.9 m/s, P = 0.04, cSBP: -8 ± 14 vs. -18 ± 18 mmHg, P < 0.01). No group differences in PWV or cSBP were observed during daytime (T2D vs. controls, PWV: 9.2 ± 1.1 vs. 9.2 ± 1.3 m/s, P = 0.99, cSBP: 133 ± 19 vs. 137 ± 25 mmHg, P = 0.42) or nighttime (PWV: 8.9 ± 1.3 vs. 8.4 ± 1.3 m/s, P = 0.14, cSBP 124 ± 20 vs. 118 ± 27 mmHg, P = 0.26). The study findings indicate that the nocturnal dipping of PWV and cSBP is attenuated in T2DM patients. The significant number of missing measurements raises concerns regarding the clinical utility of the Arteriograph 24 device.
本研究旨在评估 Arteriograph 24 设备测量 2 型糖尿病(T2DM)患者和非糖尿病对照者 24 小时脉搏波速度(PWV)和中心收缩压(cSBP)的可行性,并比较两组患者的日间和夜间特征。共测量了 58 例 T2DM 患者(平均年龄:66±9 岁,50%为女性,T2DM 病程平均:7.8±1.5 年)和 62 例年龄和性别匹配的对照者的 24 小时 PWV 和 cSBP。70%的参与者(71%的 T2DM 患者和 69%的对照者)有足够的读数生成可接受的 24 小时报告(≥14 天和≥7 个夜间读数)。两组患者的夜间 PWV 和 cSBP 均低于日间。与对照组相比,T2DM 患者的夜间 PWV 和 cSBP 下降幅度较小(PWV:-0.3±0.9 比-0.7±0.9 m/s,P=0.04,cSBP:-8±14 比-18±18 mmHg,P<0.01)。两组患者日间(T2D 与对照组,PWV:9.2±1.1 比 9.2±1.3 m/s,P=0.99,cSBP:133±19 比 137±25 mmHg,P=0.42)或夜间(PWV:8.9±1.3 比 8.4±1.3 m/s,P=0.14,cSBP:124±20 比 118±27 mmHg,P=0.26)的 PWV 或 cSBP 均无差异。研究结果表明,T2DM 患者的 PWV 和 cSBP 夜间下降幅度减弱。大量的缺失测量值引起了对 Arteriograph 24 设备临床实用性的关注。