Department of Radiology, Niigata Prefectural Central Hospital, 205, Shin-minamimachi, Niigata, 205943-0192, Japan.
Department of Radiology, Chugoku Central Hospital of the Mutual Aid Association of Public School Teachers, 148-13, Miyuki, Fukuyama, Hiroshima, 720-2121, Japan.
Phys Eng Sci Med. 2024 Jun;47(2):589-596. doi: 10.1007/s13246-024-01388-y. Epub 2024 Feb 19.
To investigate the impact of sagging correction calibration errors in radiotherapy software on image matching. Three software applications were used, with and without a polymethyl methacrylate rod supporting the ball bearings (BB). The calibration error for sagging correction across nine flex maps (FMs) was determined by shifting the BB positions along the Left-Right (LR), Gun-Target (GT), and Up-Down (UD) directions from the reference point. Lucy and pelvic phantom cone-beam computed tomography (CBCT) images underwent auto-matching after modifying each FM. Image deformation was assessed in orthogonal CBCT planes, and the correlations among BB shift magnitude, deformation vector value, and differences in auto-matching were analyzed. The average difference in analysis results among the three softwares for the Winston-Lutz test was within 0.1 mm. The determination coefficients (R) between the BB shift amount and Lucy phantom matching error in each FM were 0.99, 0.99, and 1.00 in the LR-, GT-, and UD-directions, respectively. The pelvis phantom demonstrated no cross-correlation in the GT direction during auto-matching error evaluation using each FM. The correlation coefficient (r) between the BB shift and the deformation vector value was 0.95 on average for all image planes. Slight differences were observed among software in the evaluation of the Winston-Lutz test. The sagging correction calibration error in the radiotherapy imaging system was caused by an auto-matching error of the phantom and deformation of CBCT images.
研究放射治疗软件中下垂校正校准误差对图像匹配的影响。使用三种软件应用程序,分别带有和不带有支撑球轴承 (BB) 的聚甲基丙烯酸甲酯杆。通过将 BB 位置沿左-右 (LR)、枪-靶 (GT) 和上-下 (UD) 方向从参考点偏移,确定九个 Flex 图 (FM) 中下垂校正的校准误差。修改每个 FM 后,对 Lucy 和骨盆体模锥形束计算机断层扫描 (CBCT) 图像进行自动匹配。在正交 CBCT 平面评估图像变形,并分析 BB 移位量、变形向量值和自动匹配之间差异的相关性。三种软件在 Winston-Lutz 测试中的分析结果平均差异在 0.1mm 以内。在每个 FM 中,LR、GT 和 UD 方向上 BB 移位量与 Lucy 体模匹配误差之间的决定系数 (R) 分别为 0.99、0.99 和 1.00。在使用每个 FM 评估自动匹配误差时,骨盆体模在 GT 方向上没有交叉相关性。所有图像平面的 BB 移位和变形向量值之间的相关系数 (r) 平均为 0.95。在评估 Winston-Lutz 测试时,软件之间存在细微差异。放射治疗成像系统中的下垂校正校准误差是由于体模的自动匹配误差和 CBCT 图像的变形引起的。