A multicentre study evaluating NG-Test DetecTool OXA-23 for the detection of OXA-23 carbapenemase directly from positive blood cultures (PBCs).
The NG-Test DetecTool OXA-23 is an immunoassay that integrates a sample preparation device. We evaluated NG-Test DetecTool OXA-23 on 189 spiked and 126 clinical PBCs. The clinical samples' standard-of-care procedure consisted of bacterial identification from the first day of positivity by MALDI-TOF MS, conventional culture and antimicrobial susceptibility testing. The immunoassay results were verified molecularly. The strains used for the spiked samples consisted of well-characterized and strains.
The NG-Test DetecTool OXA-23 was evaluated on 315 PBCs and revealed sensitivity of 100% (95% CI: 98.21%-100.00%) and specificity of 100% (95% CI: 96.73%-100.00%). It provided 204 true-positive results for OXA-23 in 196 bottles with carbapenem-resistant (CRAB) and 8 bottles with carbapenem-resistant and also provided 111 true-negative results. There were no false-positive and no false-negative results. Among the 315 PBCs studied, 83 clinical blood cultures collected in the ICU of a Greek university hospital, which were tested prospectively, all yielded CRAB, and OXA-23 was correctly detected in all samples from the first day of positivity using the NG-Test DetecTool OXA-23.
The NG-Test DetecTool OXA-23 has exhibited excellent sensitivity and specificity for OXA-23 detection in PBCs and can provide valuable information for appropriate selection of antibiotic therapy and early implementation of infection control measures.
一项多中心研究,评估NG-Test DetecTool OXA-23用于直接从阳性血培养物(PBC)中检测OXA-23碳青霉烯酶。
NG-Test DetecTool OXA-23是一种整合了样本制备装置的免疫测定法。我们在189份加标样本和126份临床PBC上评估了NG-Test DetecTool OXA-23。临床样本的标准护理程序包括从阳性第一天起通过基质辅助激光解吸电离飞行时间质谱(MALDI-TOF MS)进行细菌鉴定、传统培养和抗菌药敏试验。免疫测定结果通过分子方法进行验证。用于加标样本的菌株包括特征明确的菌株。
在315份PBC上对NG-Test DetecTool OXA-23进行了评估,结果显示敏感性为100%(95%置信区间:98.21%-100.00%),特异性为100%(95%置信区间:96.73%-100.00%)。它在196瓶耐碳青霉烯类鲍曼不动杆菌(CRAB)和8瓶耐碳青霉烯类其他细菌中为OXA-23提供了204个真阳性结果,还提供了111个真阴性结果。没有假阳性和假阴性结果。在所研究的315份PBC中,希腊一所大学医院重症监护病房前瞻性检测的83份临床血培养物均培养出CRAB,使用NG-Test DetecTool OXA-23从阳性第一天起在所有样本中均正确检测出OXA-23。
NG-Test DetecTool OXA-23在PBC中检测OXA-23表现出优异的敏感性和特异性,可为适当选择抗生素治疗和早期实施感染控制措施提供有价值的信息。