Abstract

PURPOSE

Measurement of cortisol concentrations is method dependent. The study aimed to establish assay-specific cut-off limits for cortisol after adrenocorticotropic hormone (ACTH) stimulation, comparing Roche Elecsys Cortisol II immunoassay to liquid chromatography-mass spectrometry (LC-MS/MS), and to assess the impact of patient characteristics, estrogen containing oral contraceptives as well as relation to other adrenocortical steroid hormone dynamics.

METHODS

One hundred healthy participants underwent a 250 μg ACTH-test, with plasma samples analyzed using ElecsysCortI, ElecsysCortII, and LC-MS/MS. Cortisone, corticosterone, 17-OH-progesterone, dehydroepiandrosterone sulfate (DHEAS), androstenedione, and testosterone were additionally analyzed with LC-MS/MS. Cut-off limit for a normal cortisol response to the ACTH-test was defined as: 2.5th percentile-1.96 × SE.

RESULTS

ElecsysCort II measured cortisol concentrations 21% (95% CI: 19-22%) lower than ElecsysCort I. Cut-off limits for cortisol 30 and 60 min after ACTH were 426 and 485 nmol/L (ElecsysCort II) and 411 and 470 nmol/L (LC-MS/MS). Cut-offs were unaffected by gender, or body-composition. The ACTH-test resulted in significantly increased adrenocortical steroid hormones, except for decreased cortisone concentrations (both sexes), and decreased testosterone in men (1.9 nmol/L, 95% CI: 1.3-2.5). Testosterone was increased in women (0.07 nmol/L, 95% CI: 0.02-0.13).

CONCLUSION

ElecsysCort II has high analytical performance and yields significantly lower cortisol concentrations than prior polyclonal immunoassays. This clinically relevant difference underscores the necessity for revised cut-off limits for improved diagnostic precision. Suggested 30-minute cortisol cutoff limits are 411 nmol/L (LC-MS/MS) and 426 nmol/L (ElecsysCort II). Adrenocortical steroids increased upon ACTH stimulation, except for cortisone in both sexes and testosterone in men, both of which decreased.

摘要

目的

皮质醇浓度的测量方法取决于方法。本研究旨在建立促肾上腺皮质激素(ACTH)刺激后皮质醇的特定检测限,比较罗氏 Elecsys Cortisol II 免疫分析法与液相色谱-质谱联用(LC-MS/MS),并评估患者特征、含雌激素的口服避孕药以及与其他肾上腺皮质类固醇激素动态的影响。

方法

100 名健康参与者接受了 250μg ACTH 测试,使用 ElecsysCortI、ElecsysCortII 和 LC-MS/MS 分析血浆样本。还使用 LC-MS/MS 分析皮质酮、皮质醇、17-羟孕酮、脱氢表雄酮硫酸酯(DHEAS)、雄烯二酮和睾酮。ACTH 测试中正常皮质醇反应的截止值定义为:第 2.5 个百分位数-1.96×SE。

结果

ElecsysCort II 测量的皮质醇浓度比 ElecsysCort I 低 21%(95%CI:19-22%)。ACTH 后 30 和 60 分钟的皮质醇截止值分别为 426 和 485nmol/L(ElecsysCort II)和 411 和 470nmol/L(LC-MS/MS)。截止值不受性别或身体成分的影响。ACTH 测试导致肾上腺皮质类固醇激素显著增加,除了皮质酮浓度降低(两性)和男性睾酮降低(1.9nmol/L,95%CI:1.3-2.5)。女性的睾酮增加(0.07nmol/L,95%CI:0.02-0.13)。

结论

ElecsysCort II 具有较高的分析性能,产生的皮质醇浓度明显低于先前的多克隆免疫分析。这种临床相关差异突出表明,为了提高诊断精度,需要修订截止值。建议的 30 分钟皮质醇截止值分别为 411nmol/L(LC-MS/MS)和 426nmol/L(ElecsysCort II)。ACTH 刺激后肾上腺皮质类固醇增加,除两性的皮质酮和男性的睾酮降低外,后者降低。