Agreement and internal quality assurance of the Neubauer hemocytometer and Makler chamber for human sperm concentration determination.

机构信息

Department of Clinical Analysis, Federal University of Santa Catarina, Florianópolis, Brazil.

出版信息

JBRA Assist Reprod. 2024 Jun 1;28(2):215-223. doi: 10.5935/1518-0557.20240023.

Abstract

OBJECTIVE

The Neubauer hemocytometer, as well as the Makler chamber, are devices commonly used in andrology laboratories. The present study aimed to verify if both methods yield comparable results, and whether they can be used interchangeably to determine sperm concentration.

METHODS

Sperm and latex beads concentration measurements were performed with the Neubauer hemocytometer and the Makler chamber. Fixed and proportional biases were estimated, and the method agreement was determined by assessing sperm concentration results with the Bland and Altman plot. The Coefficient of Variation (CV) and relative bias were calculated as an index of precision and accuracy, respectively, by measuring latex beads target concentrations in both chambers.

RESULTS

The Makler chamber systematically overestimated the Neubauer hemocytometer concentration measurements by a mean of -7.99%, with limits of agreement (LOA) between -41% to 25.61% (p<0.001). The fixed bias was found for concentration values inferior to 40 x 106/ml range (p<0.001), but not higher concentration results (p>0.05). Measurements with the Neubauer hemocytometer showed the greatest consistency in the study with the CV ranging from 3.01% to 6.67%; while the CV with the Makler chamber ranged from 8.46% to 25.64%. The relative bias for the Neubauer hemocytometer determinations varied from 0.12% to 8.40%, while for the Makler chamber varied from 7.6% to an overestimation of 38.0%.

CONCLUSIONS

Measurements made with the Makler chamber demonstrated more variability and a higher degree of overestimation. The Makler chamber is a poor substitute to the Neubauer hemocytometer for evaluation of oligozoospermic samples, although both chambers render similar results for highly concentrated samples.

摘要

目的

新伯氏血细胞计数板和迈勒腔都是在男科实验室中常用的设备。本研究旨在验证这两种方法是否能得到可比的结果,以及它们是否可以互换使用以确定精子浓度。

方法

使用新伯氏血细胞计数板和迈勒腔测量精子和乳胶珠的浓度。估计固定和比例偏差,并通过 Bland 和 Altman 图评估精子浓度结果来确定方法的一致性。通过在两个腔室中测量乳胶珠目标浓度来计算变异系数(CV)和相对偏差,分别作为精密度和准确性的指标。

结果

迈勒腔系统地高估了新伯氏血细胞计数板浓度测量值,平均低估了-7.99%,其一致性范围为-41%至 25.61%(p<0.001)。固定偏差仅存在于浓度值低于 40 x 106/ml 范围(p<0.001),而不存在于更高浓度结果中(p>0.05)。新伯氏血细胞计数板的测量值在研究中表现出最大的一致性,CV 范围为 3.01%至 6.67%;而迈勒腔的 CV 范围为 8.46%至 25.64%。新伯氏血细胞计数板测定的相对偏差范围为 0.12%至 8.40%,而迈勒腔的相对偏差范围为 7.6%至 38.0%的高估。

结论

迈勒腔的测量值表现出更大的变异性和更高的高估程度。虽然对于高浓度样本,两个腔室都能得到相似的结果,但迈勒腔不能替代新伯氏血细胞计数板来评估少精子症样本。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/74b0/11152427/20a0c10eaace/jbra-28-02-0215-g01.jpg