通过对拍摄的微珠进行精确计数来评估不同精子计数室的准确性。

Evaluating the accuracy of different sperm counting chambers by performing strict counts of photographed beads.

作者信息

Shiran E, Stoller J, Blumenfeld Z, Feigin P D, Makler A

机构信息

Department of Obstetrics and Gynecology, Rambam Medical Center, Haifa, Israel.

出版信息

J Assist Reprod Genet. 1995 Aug;12(7):434-42. doi: 10.1007/BF02211144.

Abstract

PURPOSE

To suggest a method for evaluating the accuracy of different kinds of sperm counting chambers by eliminating errors concerned with human skills or semen properties. In this method, various concentrations were prepared from stocks of commercially available latex beads (Accu-beads, Hamilton-Thorn Research, Beverly, MA). Samples from identical preparations were loaded into different types of chambers, namely, hemocytometer (Neubauer, improved double, Superior Ltd, Germany), Makler (Sefi Medical Instruments, Haifa, Israel) and Horwell (Horwell Ltd, London, UK). Beads were counted by both direct microscopic observations and by strict scanning of their photographed images.

RESULTS

In all cases, counts by direct observation were about 5% higher than strict counts of the same photographed beads. Counting photographed beads showed high reproducibility (average CV of 5.1%) between samples in the two wells of the hemocytometer. Counts of photographed beads, sampled from identical stocks, were on average slightly lower in the Makler chamber (20.7 x 10(6)/ml) and much higher in the Horwell chamber (47.4 x 10(6)/ml) than counts in the hemocytometer (21.5 x 10(6)/ml). Samples from three different batches of Accubeads revealed slight variation in counts between the batches and an average concentration of 11% above the number indicated on the commercial product. CONCLUSIONS; A technique that combined loading latex beads from identical stock into various chambers, proper covering of the tested samples and strict counting of photographed beads provided precise and reproducible results. By eliminating most errors related to human skills and semen properties, this method is suitable for evaluating the accuracy of counting chambers.

摘要

目的

提出一种评估不同类型精子计数室准确性的方法,以消除与人为操作技能或精液特性相关的误差。在该方法中,从市售乳胶珠(Accu-beads,Hamilton-Thorn Research,马萨诸塞州贝弗利)储备液中制备不同浓度的样本。将相同制剂的样本加载到不同类型的计数室中,即血细胞计数板(Neubauer改良双板,Superior Ltd,德国)、Makler计数板(Sefi Medical Instruments,海法,以色列)和Horwell计数板(Horwell Ltd,伦敦,英国)。通过直接显微镜观察和对拍摄图像的严格扫描来计数乳胶珠。

结果

在所有情况下,直接观察计数比相同拍摄乳胶珠的严格计数高约5%。血细胞计数板两孔样本间拍摄乳胶珠计数显示出高重复性(平均变异系数为5.1%)。从相同储备液中取样的拍摄乳胶珠计数,在Makler计数板中平均略低(20.7×10⁶/ml),在Horwell计数板中则比血细胞计数板(21.5×10⁶/ml)高得多(47.4×10⁶/ml)。来自三批不同Accubeads的样本显示批次间计数略有差异,平均浓度比商业产品标注的数值高11%。结论:将相同储备液中的乳胶珠加载到不同计数室、对测试样本进行适当覆盖以及对拍摄乳胶珠进行严格计数的技术可提供精确且可重复的结果。通过消除大多数与人为操作技能和精液特性相关的误差,该方法适用于评估计数室的准确性。