Houk T L, Kruger R A, Mistretta C A, Riederer S J, Shaw C G, Lancaster J C, Flemming D C
Invest Radiol. 1979 Jul-Aug;14(4):270-8. doi: 10.1097/00004424-197907000-00002.
We report in vitro and in vivo trials of K-edge fluoroscopy, by which iodine contrast concentration is displayed live, with tissue and bone images suppressed, free of patient-motion artifacts. Iodine and cerium, 125 and 225 mg/cm2 respectively, filter alternate TV fields of cine-pulsed 50 KVP x-rays. Weighted subtraction of successive TV fields isolates the iodine image and simultaneously minimizes artifacts. Digital techniques are used in real time. At our present x-ray tube limit, 500 mA instantaneous current, the patient exposure is 180 mR/sec and quantum mottle limits the image quality. Integrating four successive difference images provides a compromise between mottle and smoothly moving displays. Cardiovascular images of a 17-kg dog, using 1 ml/kg Renografin-60 injected into a foreleg vein, show that a 15-cm chest thickness is our present practical maximum. This method may be useful in diagnosing cardiovascular anomalies in infants without catheterization or suspension of breathing.
我们报告了K边荧光透视的体外和体内试验,通过该试验可实时显示碘造影剂浓度,同时抑制组织和骨骼图像,且无患者运动伪影。碘和铈分别以125和225mg/cm²的厚度过滤电影脉冲50KVP X射线的交替电视场。对连续电视场进行加权相减可分离出碘图像,同时将伪影降至最低。实时使用数字技术。在我们目前的X射线管极限(500mA瞬时电流)下,患者的辐射剂量为180mR/秒,量子斑点限制了图像质量。整合四个连续的差分图像可在斑点和平滑移动显示之间取得折衷。对一只17kg的狗进行心血管成像,通过将1ml/kg的泛影葡胺-60注入前肢静脉,结果表明目前实际可行的最大胸部厚度为15cm。该方法可能有助于在不进行导管插入或暂停呼吸的情况下诊断婴儿心血管异常。