Biomedical Imaging Research Group (GIBI2(30)), La Fe Health Research Institute (IIS La Fe), Avenida Fernando Abril Martorell, València 46026, Spain.
Biomedical Imaging Research Group (GIBI2(30)), La Fe Health Research Institute (IIS La Fe), Avenida Fernando Abril Martorell, València 46026, Spain; Nuclear Medicine Department, La Fe University and Polytechnic Hospital, Avenida Fernando Abril Martorell, València 46026, Spain.
Phys Med. 2024 May;121:103345. doi: 10.1016/j.ejmp.2024.103345. Epub 2024 Apr 5.
To evaluate whether the Centiloid Scale may be used to diagnose Alzheimer's Disease (AD) pathology effectively with the only use of amyloid PET imaging modality from a brain-dedicated PET scanner.
This study included 26 patients with amyloid PET images with 3 different radiotracers. All patients were acquired both on a PET/CT and a brain-dedicated PET scanner (CareMiBrain, CMB), from which 4 different reconstructions were implemented. A new pipeline was proposed and used for the PET image analysis based on the original Centiloid Scale processing pipeline, but with only PET images. The Youden's Index was employed to calculate the optimal cutoffs for diagnosis and evaluated by the AUC, accuracy, precision, and recall metrics.
The Centiloid Scale (CL) processing pipeline was validated with and without the use of MR images. The CL cutoffs for AD pathology diagnosis on the PET/CT and the 4 CMB reconstructions were 34.4 ± 2.2, 43.5 ± 3.5, 51.9 ± 12.5, 57.5 ± 6.8 and 41.8 ± 1.2 respectively. Overall, for these cutoffs all metrics obtained the maximum score.
The Centiloid scale applied to PET images allows for AD pathology diagnosis. The CMB scanner can be used with the Centiloid scale to automatically assist in the diagnosis of AD pathology, relieving the large burden of neurodegenerative diseases on a traditional PET/CT.
评估 Centiloid 量表是否可以仅使用脑专用 PET 扫描仪的淀粉样蛋白 PET 成像模式有效地诊断阿尔茨海默病(AD)病理。
本研究纳入了 26 名具有 3 种不同放射性示踪剂的淀粉样蛋白 PET 图像患者。所有患者均在 PET/CT 和脑专用 PET 扫描仪(CareMiBrain,CMB)上采集,从这两种设备中实施了 4 种不同的重建。提出了一种新的流水线,并基于原始 Centiloid 量表处理流水线但仅使用 PET 图像进行 PET 图像分析。使用 Youden's 指数计算诊断的最佳截断值,并通过 AUC、准确性、精度和召回率进行评估。
验证了在使用和不使用 MR 图像的情况下 Centiloid 量表(CL)处理流水线。在 PET/CT 和 4 种 CMB 重建中,AD 病理诊断的 CL 截断值分别为 34.4±2.2、43.5±3.5、51.9±12.5、57.5±6.8 和 41.8±1.2。总体而言,对于这些截断值,所有指标均获得最高分。
应用于 PET 图像的 Centiloid 量表允许 AD 病理诊断。CMB 扫描仪可以与 Centiloid 量表一起使用,以自动协助 AD 病理诊断,减轻传统 PET/CT 上神经退行性疾病的巨大负担。