Private Ophthalmology Practice, Atieh Medical Center, Rasht, 4144694198, Guilan, Iran.
Department of Ophthalmology and Visual Science, Gavin Herbert Eye Institute, University of California at Irvine, Irvine, CA, USA.
BMC Ophthalmol. 2024 Apr 19;24(1):179. doi: 10.1186/s12886-024-03459-w.
Plusoptix photoscreeners are capable of measuring refractive errors of children from 1 meter distance, without cyloplegia. We aimed to compare refractive data obtained from the newest version of Plusoptix (model 12) with cycloplegic autorefraction.
We examined 111 consecutive children aged 3-7 years first by Plusoptix A12C under manifest condition and subsequently for cycloplegic refraction by Topcon KR-1 tabletop autorefractometer. Sphere, spherical equivalent, cylinder and axis of astigmatism measured by the two methods were analyzed to determine correlation, agreement and differences.
Binocular examination of 111 children aged 4.86±1.27 years revealed good agreement between refractive data obtained by Plusoptix and cycloautorefraction, according to Bland-Altman plots. Significant (p < 0.001) and strong correlation was found between all refractive measurements (Pearson's r value of 0.707 for sphere, 0.756 for pherical equivalent, and 0.863 for cylinder). Plusoptix mean sphere, spherical equivalent and cylinder were 1.22, 0.56, and -1.32 D, respectively. Corresponding values for cycloautorefraction were 1.63, 1.00, and -1.26 D. The difference between axis of cylinder measured by the two methods was < 10° in 144 eyes (64.9%).
Considering the significant agreement and correlation between Plusoptix photoscreener and cycloplegic autorefraction, the need for cycloplegic drops in refractive examination of children may be obviated. The mean difference between cylinder measurements are considerably trivial (0.06 D), but sphere is approximately 0.4 D underestimated by Plusoptix compared to cycloautorefraction, on average.
Plusoptix 视力筛查仪能够在无需睫状肌麻痹的情况下,从 1 米的距离测量儿童的屈光不正。我们旨在比较新型 Plusoptix(型号 12)与睫状肌麻痹自动验光仪获得的屈光数据。
我们使用 Plusoptix A12C 在明视状态下对 111 例年龄在 3-7 岁的连续儿童进行了首次检查,随后使用 Topcon KR-1 台式自动验光仪进行睫状肌麻痹验光。分析两种方法测量的球镜、等效球镜、柱镜和散光轴位,以确定相关性、一致性和差异。
对 111 例年龄为 4.86±1.27 岁的儿童进行双眼检查,根据 Bland-Altman 图,Plusoptix 和睫状肌麻痹自动验光仪获得的屈光数据之间具有良好的一致性。所有屈光测量均存在显著(p < 0.001)和强相关性(球镜为 Pearson's r 值为 0.707,等效球镜为 0.756,柱镜为 0.863)。Plusoptix 平均球镜、等效球镜和柱镜分别为 1.22、0.56 和-1.32 D,睫状肌麻痹自动验光仪的相应值为 1.63、1.00 和-1.26 D。两种方法测量的柱镜轴位差值<10°的眼数为 144 只(64.9%)。
考虑到 Plusoptix 视力筛查仪与睫状肌麻痹自动验光之间具有显著的一致性和相关性,儿童屈光检查可能无需使用睫状肌麻痹剂。两种方法测量的柱镜差值相当小(0.06 D),但平均而言,Plusoptix 测量的球镜值比睫状肌麻痹自动验光仪低约 0.4 D。