William Pritchard Veterinary Medical Teaching Hospital, School of Veterinary Medicine, University of California Davis, Davis, CA, USA.
Department of Surgical and Radiological Sciences, School of Veterinary Medicine, University of California Davis, Davis, CA, USA.
Vet Anaesth Analg. 2024 Jul-Aug;51(4):322-332. doi: 10.1016/j.vaa.2024.03.003. Epub 2024 Mar 9.
To evaluate the effect of increased respiratory system resistance (R) on dynamic compliance (C) assessed by the NM3 monitor (C) and the E-CAiOV module (C).
Prospective laboratory study.
A training test lung (TTL) simulated the mechanical ventilation of a mammal with 50 and 300 mL tidal volumes in three conditions of R [normal (R), moderately increased (R) and severely increased (R)] and a wide range of clinically relevant C. Simulations at increased R were paired with simulations at R with the same static compliance for comparisons. Pearson's correlation coefficient and concordance correlation coefficient between the measurements at R with the ones with increased R were calculated. Bland-Altman plots were also used to evaluate the agreement of C and C at R (control values) with their paired values at R and R. Relative bias and limits of agreement (LOAs) were calculated and LOAs larger than 30% were considered unacceptable. Trending ability of C and C were evaluated by polar plots. Values of p < 0.05 were considered significant.
The effect of increased R was more pronounced for C than for C. Unacceptable agreement was only observed in C at R in the 300 mL simulation (bias = -18.3% and lower LOA = -45%). For C, agreement was unacceptable for all tested R in both simulations, being the worst at R in the 300 mL simulation (bias = -54.7% and lower LOA = -100.2%). Both levels of increased R caused poor trending ability for C, whereas the same effect was only observed for C at R.
In the presence of increased R, C estimated by the NM3 monitor presented better capability to distinguish between changes in R from changes in respiratory system compliance.
评估呼吸系统阻力(R)增加对 NM3 监测仪(C)和 E-CAiOV 模块(C)评估的动态顺应性(C)的影响。
前瞻性实验室研究。
一个训练测试肺(TTL)模拟了哺乳动物的机械通气,在 R [正常(R)、中度增加(R)和严重增加(R)]和广泛的临床相关 C 三种条件下,具有 50 和 300 毫升潮气量。对增加 R 的模拟与具有相同静态顺应性的 R 模拟进行配对比较。计算 R 与增加 R 时测量值之间的 Pearson 相关系数和一致性相关系数。还使用 Bland-Altman 图评估 C 和 C 在 R(对照值)与其在 R 和 R 时的配对值的一致性。计算相对偏差和一致性界限(LOA),并认为 LOA 大于 30%为不可接受。通过极坐标图评估 C 和 C 的趋势能力。p 值小于 0.05 被认为具有统计学意义。
增加 R 对 C 的影响比对 C 的影响更为显著。仅在 300 毫升模拟的 R 时观察到 C 的一致性不可接受(偏差=-18.3%,较低的 LOA=-45%)。对于 C,在两种模拟中,所有测试的 R 均不一致,在 300 毫升模拟的 R 时最差(偏差=-54.7%,较低的 LOA=-100.2%)。增加 R 的两个水平都导致 C 的趋势能力较差,而 C 仅在 R 时出现这种情况。
在 R 增加的情况下,NM3 监测仪估计的 C 能够更好地区分 R 变化与呼吸系统顺应性变化。