A Comprehensive Study on Implant Stability Quotient (ISQ) in View of Resonance Frequency and Spectrum Analysis.

出版信息

Int J Oral Maxillofac Implants. 2024 Aug 29;39(4):567-574. doi: 10.11607/jomi.10606.

Abstract

PURPOSE

To investigate how well an implant stability quotient (ISQ) represents resonance frequency.

MATERIALS AND METHODS

Benchtop experiments on standardized samples that replicated a mandibular premolar site were conducted to correlate an ISQ value and a resonance frequency to synthetic bone density and an incremental insertion torque; then, a frequency spectrum analysis was performed to check the validity of the resonance frequency analysis (RFA). Brånemark Mk III implants (4 × 11.5 mm; Nobel Biocare) were placed in Sawbones test models of five different densities (40, 30, 40/20, 20, and 15 PCF). An incremental insertion torque was recorded during implant placement. To perform stability measurements, the test models were partially clamped in a vise (unclamped volume: 10 × 20 × 34 mm). A MulTipeg (Integration Diagnostics) was attached to the implants, and a Penguin (Integration Diagnostics) RFA measured ISQ. Simultaneously, the MulTipeg motion was monitored via a laser Doppler vibrometer and processed by a spectrum analyzer to obtain the resonance frequency. Tightness of the clamp was adjusted to vary the resonance frequency. A statistical analysis produced a linear correlation coefficient (R) among the measured ISQ, resonance frequency, and incremental insertion torque.

RESULTS

The resonance frequency had high correlation to the incremental insertion torque (R = 0.978), confirming the validity of using RFA for this study. Measured ISQ data were scattered and had low correlation to the resonance frequency (R = 0.214) and the incremental insertion torque (R = 0.386). The spectrum analysis revealed the simultaneous presence of multiple resonance frequencies.

CONCLUSIONS

For the designed benchtop tests, resonance frequency does indicate implant stability in view of Sawbones density and incremental insertion torque. However, ISQ measurements do not correlate to the resonance frequency and may not reflect the stability when multiple resonance frequencies are present simultaneously.

摘要

目的

研究种植体稳定系数(ISQ)与共振频率的相关性。

材料与方法

在标准化样本上进行台架实验,以模拟下颌前磨牙部位,将 ISQ 值与共振频率与合成骨密度和递增插入扭矩相关联;然后,进行频谱分析以检查共振频率分析(RFA)的有效性。将 Brånemark Mk III 种植体(4×11.5mm;诺保科)植入五种不同密度(40、30、40/20、20 和 15PCF)的 Sawbones 测试模型中。在植入过程中记录递增插入扭矩。为了进行稳定性测量,将测试模型部分夹在虎钳中(未夹紧体积:10×20×34mm)。将 MulTipeg(集成诊断)连接到种植体上,Penguin(集成诊断)RFA 测量 ISQ。同时,通过激光多普勒测振仪监测 MulTipeg 运动,并通过频谱分析仪进行处理以获得共振频率。调整虎钳的紧固度以改变共振频率。统计分析产生了测量的 ISQ、共振频率和递增插入扭矩之间的线性相关系数(R)。

结果

共振频率与递增插入扭矩高度相关(R=0.978),证实了使用 RFA 进行本研究的有效性。测量的 ISQ 数据分散,与共振频率(R=0.214)和递增插入扭矩(R=0.386)相关性低。频谱分析显示同时存在多个共振频率。

结论

就设计的台架测试而言,考虑到 Sawbones 密度和递增插入扭矩,共振频率确实表明种植体的稳定性。然而,ISQ 测量与共振频率不相关,并且当同时存在多个共振频率时,可能无法反映稳定性。