Department of Oral Surgery, Faculty of Dental Medicine, Saint Joseph University, Beirut, Lebanon.
Implant Dent. 2012 Dec;21(6):474-80. doi: 10.1097/ID.0b013e31826918f1.
To determine if resonance frequency analysis (RFA) correlated with insertion torque (IT), bone volume (BV), and clinical osseointegration at 6 weeks.
For 18 patients, BV was evaluated histologically by retrieving bone core biopsies before placement of 40 dental implants. Peak IT was recorded at implant placement, and RFA values (implant stability quotient [ISQ]) were noted at baseline, 3 weeks, and 6 weeks. Osseointegration was evaluated at 6 weeks when torquing abutments to 30 N. ISQ values were correlated with IT, BV and abutment torquing results. Data were statistically analyzed.
ISQ values significantly decreased at 3 weeks and increased at 6 weeks. There was a significant positive correlation between BV and ISQ at baseline and at 3 weeks, but not at 6 weeks, and between ISQ and jaw location, implant diameter and IT at baseline, 3 weeks, and 6 weeks. There was a significant correlation between spinning/painful implants during abutment torquing and low ISQ, low BV, and low IT values.
Correlations between BV and IT values, and ISQ suggest that RFA may indicate primary implant stability. BV, IT, and ISQ values may anticipate the degree of osseointegration at 6 weeks.
确定共振频率分析(RFA)是否与植入扭矩(IT)、骨量(BV)和 6 周时的临床骨整合相关。
在放置 40 颗牙科种植体之前,对 18 名患者通过取出骨芯活检来评估 BV。在植入时记录峰值 IT,在基线、3 周和 6 周时记录 RFA 值(种植体稳定性指数 [ISQ])。当将基台拧至 30 N 时,在 6 周时评估骨整合。ISQ 值在 3 周时显著降低,在 6 周时增加。在基线和 3 周时,BV 与 ISQ 呈显著正相关,但在 6 周时没有,在基线、3 周和 6 周时,ISQ 与颌骨位置、种植体直径和 IT 呈显著正相关。在基台拧紧时旋转/疼痛的种植体与低 ISQ、低 BV 和低 IT 值之间存在显著相关性。
BV 和 IT 值与 ISQ 之间的相关性表明 RFA 可能表明初始种植体稳定性。BV、IT 和 ISQ 值可能预测 6 周时的骨整合程度。