Suppr超能文献

在犬类动物中验证一种快速数字血管造影技术,用于在常规心脏导管插入术期间测量相对冠状动脉血流量。

Validation in dogs of a rapid digital angiographic technique to measure relative coronary blood flow during routine cardiac catheterization.

作者信息

Hodgson J M, LeGrand V, Bates E R, Mancini G B, Aueron F M, O'Neill W W, Simon S B, Beauman G J, LeFree M T, Vogel R A

出版信息

Am J Cardiol. 1985 Jan 1;55(1):188-93. doi: 10.1016/0002-9149(85)90326-1.

Abstract

Assessment of the functional significance of anatomically defined coronary stenoses has been hampered by the lack of clinically applicable techniques of measuring coronary blood flow or flow ratios. A digital angiographic technique is reported that allows rapid analysis of relative regional coronary blood flow during routine cardiac catheterization. This technique was validated in dogs by comparing digital flow ratio estimates with electromagnetic-flow (EMF) ratio measurements. Fourteen open-chest dogs had EMF probes placed on the proximal left anterior descending artery before selective coronary angiography. Electrocardiographically gated images were acquired directly by a digital radiographic system during both baseline blood flow and either contrast or papaverine-induced hyperemia. Dual-parameter functional images were generated using color and intensity coding to represent contrast arrival time and contrast density, respectively. For analysis, myocardial areas of interest were created over the distal perfusion bed of the left anterior descending coronary artery. Mean contrast density/appearance time (CD/AT) values were computer calculated as the mean density divided by the mean arrival time for each. Coronary flow reserve was determined as the ratio of the CD/AT value for a hyperemic image divided by the CD/AT value for the corresponding baseline image. CD/AT ratios correlated well (r = 0.92) with actual EMF ratios (CD/AT Ratio = 0.90 EMF Ratio +0.12, n = 48 ratios). Reproducibility was +/- 13%. Interobserver (r = 0.99) and intraobserver (r = 0.98) variability was excellent. Thus, rapid, accurate and reproducible estimates of relative regional coronary blood flow are possible using digital radiography.

摘要

由于缺乏可临床应用的测量冠状动脉血流或血流比值的技术,对解剖学定义的冠状动脉狭窄的功能意义评估一直受到阻碍。本文报道了一种数字血管造影技术,该技术可在常规心脏导管插入术期间快速分析相对局部冠状动脉血流。通过将数字血流比值估计值与电磁血流(EMF)比值测量值进行比较,在犬类中验证了该技术。在选择性冠状动脉造影之前,对14只开胸犬在左前降支近端放置EMF探头。在基线血流以及造影剂或罂粟碱诱导的充血期间,通过数字射线摄影系统直接采集心电图门控图像。使用颜色和强度编码分别生成双参数功能图像,以表示造影剂到达时间和造影剂密度。为了进行分析,在左前降支冠状动脉远端灌注床上方创建心肌感兴趣区域。计算机计算平均造影剂密度/出现时间(CD/AT)值,即每个区域的平均密度除以平均到达时间。冠状动脉血流储备定义为充血图像的CD/AT值除以相应基线图像的CD/AT值。CD/AT比值与实际EMF比值相关性良好(r = 0.92)(CD/AT比值 = 0.90 EMF比值 + 0.12,n = 48个比值)。可重复性为+/- 13%。观察者间(r = 0.99)和观察者内(r = 0.98)的变异性极佳。因此,使用数字射线摄影可以快速、准确且可重复地估计相对局部冠状动脉血流。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验