Peters Frank B, Rapp Andreas O
Institut für Berufswissenschaften im Bauwesen, Leibniz-Universität Hannover, Herrenhäuser Straße 8, 30419 Hannover, Germany.
Methods Appl Fluoresc. 2024 Jun 17;12(3). doi: 10.1088/2050-6120/ad5415.
This technical note presents a device to diminish scattering signal in front-face fluorescence spectra while obtaining fluorescence signal. The beam path in a commercial fluorescence spectrometer was modified by two deflecting mirrors, leading reflections away from the sensor. This light path modifying (LPM) device was tested with two fluid and three solid substances, where the scattering-to-fluorescence ratio improved by a factor of 1.7 to 7.6. The spectra obtained with the LPM were much clearer, and distortion of the fluorescence peaks was avoided. Scans of quinine sulphate complied well with reference spectra.
本技术说明介绍了一种在获取荧光信号的同时减少正面荧光光谱中散射信号的装置。商用荧光光谱仪中的光路通过两个偏转镜进行了修改,使反射光远离传感器。这种光路修改(LPM)装置用两种流体和三种固体物质进行了测试,散射与荧光的比率提高了1.7至7.6倍。使用LPM获得的光谱要清晰得多,并且避免了荧光峰的畸变。硫酸奎宁的扫描结果与参考光谱非常吻合。