白人成年人呼吸内震荡测量参考方程。

Reference Equations for Within-Breath Respiratory Oscillometry in White Adults.

机构信息

Department of Electronics, Information and Bioengineering, Politecnico di Milano University, Milan, Italy,

RESTECH Srl, Milan, Italy.

出版信息

Respiration. 2024;103(9):521-534. doi: 10.1159/000539532. Epub 2024 Jun 7.

Abstract

BACKGROUND

Within-breath analysis of oscillometry parameters is a growing research area since it increases sensitivity and specificity to respiratory pathologies and conditions. However, reference equations for these parameters in White adults are lacking and devices using multiple sinusoids or pseudorandom forcing stimuli have been underrepresented in previous studies deriving reference equations. The current study aimed to establish reference ranges for oscillometry parameters, including also the within-breath ones in White adults using multi-sinusoidal oscillations.

METHODS

White adults with normal spirometry, BMI ≤30 kg/m2, without a smoking history, respiratory symptoms, pulmonary or cardiac disease, neurological or neuromuscular disorders, and respiratory tract infections in the previous 4 weeks were eligible for the study. Study subjects underwent oscillometry (multifrequency waveform at 5-11-19 Hz, Resmon PRO FULL, RESTECH Srl, Italy) in 5 centers in Europe and the USA according to international standards. The within-breath and total resistance (R) and reactance (X), the resonance frequency, the area under the X curve, the frequency dependence of R (R5-19), and within-breath changes of X (ΔX) were submitted to lambda-mu-sigma models for deriving reference equations. For each output parameter, an AIC-based stepwise input variable selection procedure was applied.

RESULTS

A total of 144 subjects (age 20.8-86.3 years; height 146-193 cm; BMI 17.42-29.98 kg/m2; 56% females) were included. We derived reference equations for 29 oscillatory parameters. Predicted values for inspiratory and expiratory parameters were similar, while differences were observed for their limits of normality.

CONCLUSIONS

We derived reference equations with narrow confidence intervals for within-breath and whole-breath oscillatory parameters for White adults.

摘要

背景

呼吸内分析振荡测量参数是一个日益发展的研究领域,因为它提高了对呼吸病理和状况的敏感性和特异性。然而,白种成年人的这些参数的参考方程是缺乏的,并且在以前的研究中,使用多个正弦波或伪随机强迫刺激的设备在推导参考方程方面代表性不足。本研究旨在建立白种成年人的振荡测量参数的参考范围,包括使用多正弦波的呼吸内参数。

方法

符合以下标准的白种成年人有资格参加这项研究:正常的肺量计检查,BMI≤30kg/m2,没有吸烟史、呼吸症状、肺部或心脏疾病、神经或神经肌肉疾病以及在过去 4 周内的呼吸道感染。研究对象根据国际标准在欧洲和美国的 5 个中心接受了振荡测量(多频波,5-11-19Hz,Resmon PRO FULL,RESTECH Srl,意大利)。呼吸内和总阻力(R)和电抗(X)、共振频率、X 曲线下面积、R 的频率依赖性(R5-19)以及呼吸内的 X 变化(ΔX)被提交给 lambda-mu-sigma 模型以推导出参考方程。对于每个输出参数,应用基于 AIC 的逐步输入变量选择程序。

结果

共纳入 144 名受试者(年龄 20.8-86.3 岁;身高 146-193cm;BMI 17.42-29.98kg/m2;女性占 56%)。我们推导出了 29 个振荡参数的参考方程。吸气和呼气参数的预测值相似,而其正常范围的限制则存在差异。

结论

我们为白种成年人推导了呼吸内和整个呼吸振荡参数的参考方程,置信区间较窄。