Krogsaa B, Fledelius H, Larsen J, Lund-Andersen H
Graefes Arch Clin Exp Ophthalmol. 1985;222(4-5):212-4. doi: 10.1007/BF02133677.
The optics of slit-lamp fluorophotometry were analyzed. This analysis forms the basis of the formulation of a slit-lamp technique called photometric oculometry, which makes it possible to measure intraocular distances. A comparison between ultrasonography and photometric oculometry in the determination of the ocular axial length was carried out. Measurements in 11 eyes showed statistical agreement, with a mean of 22.90 mm, SD +/- 1.815 (ultrasound), and a mean of 23.17 mm, SD +/- 1.667 (photometry). The principle involved in photometric oculometry is of special importance for the calculation of the blood-retinal barrier permeability to fluorescein by vitreous fluorophotometry, since the method allows the volume of the vitreous body to be estimated directly from the slit-lamp fluorophotometer recording.