Med Sci Monit. 2024 Jun 10;30:e944502. doi: 10.12659/MSM.944502.
BACKGROUND Before insertion, chairside adjustment kits are heat sterilized for positioning and polishing dental restorations. This study aimed to evaluate the effects of 2 steam sterilization cycles on the efficacy of polishing highly translucent monolithic zirconia (HTMLZ) dental restoration material. MATERIAL AND METHODS 100 HTMLZ disc-shaped specimens were adjusted (grinding, finishing, polishing) with EVE Diacera kit. Two steam sterilization techniques [standard (Gp S), immediate/flash (Gp (F)] of CAK were further subgrouped based on number of sterilization cycles [cycle 1 (control), cycle 5, 10, 15, and 20 (experimental)] (n=10 each). Each subgroup accordingly was evaluated for average surface roughness (Ra) and root mean square roughness (Rq) using a profilometer. Mean and standard deviation of 5 subgroups were statistically analyzed using one-way ANOVA/post hoc Tukey's test. Scanning electron microscopy complemented Ra, Rq measurements. Statistical differences of P≤0.05 were considered significant. RESULTS HTMLZ specimens in both groups showed increased (Ra/Rq) values after repeated sterilization of EVE Diacera kit, with Gp F showing lesser increase than Gp S (20 cycles). Gp F at 10 cycles and Gp S at 15 cycles showed clinically unacceptable roughness threshold (0.25 μm). Differences between subgroups for Ra and Rq values were significant (P≤0.05) with less differences within groups observed in early cycles (1, 10). Results validate the manufacturer's recommendations of using flash sterilization/10 cycles for EVE Diacera kit. CONCLUSIONS Repeated sterilization reduces efficacy of chairside adjustment kit to produce smooth surfaces on HTMLZ. This study recommends flash sterilization to a maximum of 10 times to get the clinically acceptable results of Ra and Rq.
在插入之前,椅旁调整套件经过热消毒以定位和抛光牙科修复体。本研究旨在评估两次蒸汽灭菌循环对高度半透明整体氧化锆(HTMLZ)牙科修复材料抛光效果的影响。
100 个 HTMLZ 盘状样本使用 EVE Diacera 套件进行调整(研磨、精整、抛光)。根据灭菌循环次数[循环 1(对照)、循环 5、10、15 和 20(实验)],进一步将 CAK 的两种蒸汽灭菌技术[标准(Gp S)、立即/闪蒸(Gp (F)]分为亚组(每组 10 个)。使用轮廓仪评估每个亚组的平均表面粗糙度(Ra)和均方根粗糙度(Rq)。使用单向方差分析/事后 Tukey 检验对 5 个亚组的平均值和标准差进行统计学分析。扫描电子显微镜补充了 Ra 和 Rq 的测量。P≤0.05 的差异被认为具有统计学意义。
两组中的 HTMLZ 样本在 EVE Diacera 套件重复灭菌后表现出(Ra/Rq)值增加,Gp F 比 Gp S 显示出更小的增加(20 个循环)。Gp F 在 10 个循环和 Gp S 在 15 个循环时表现出临床不可接受的粗糙度阈值(0.25 μm)。Ra 和 Rq 值的亚组间差异具有统计学意义(P≤0.05),在早期循环(1、10)中观察到组内差异较小。结果验证了制造商对 EVE Diacera 套件使用闪蒸灭菌/10 个循环的建议。
重复灭菌会降低椅旁调整套件在 HTMLZ 上产生光滑表面的效果。本研究建议闪蒸灭菌最多 10 次,以获得 Ra 和 Rq 的临床可接受结果。