荧光法在Sysmex XN9000血液分析仪上用于校正小红细胞症个体血小板计数的应用。

Application of the Fluorescence Method on Sysmex XN9000 Hematology Analyzer for Correcting Platelet Count in Individuals with Microcytosis.

作者信息

Deng Jiankai, Xie Shuhua, Chen Yaoming, Ma Qinghua, He Yuting, Liu Min, Wang Dong, Yu Xuegao

机构信息

Department of Laboratory Medicine, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, China.

出版信息

Lab Med. 2023 Jan 5;54(1):e10-e13. doi: 10.1093/labmed/lmac063.

Abstract

OBJECTIVE

Although small red blood cells are a well-known analytical pitfall that could cause artifactual increase of the platelet count, limited information is available on the accuracy of impedance platelet counting in cases with microcytosis. The aim of this study is to assess the accuracy of impedance platelet counting in the presence of small red blood cells, and to establish the optimal mean corpuscular volume (MCV) cutoff to endorse fluorescence platelet counting.

METHODS

In this study, platelet counts estimated by the impedance method on the Sysmex XN9000 analyzer (Sysmex, Kobe, Japan) were compared with those provided by the fluorescence method. The accuracy of impedance platelet counting was assessed. Receiver operating characteristic curve was used to evaluate the performance of MCV in predicting falsely increased platelet counts.

RESULTS

There was a tendency for the impedance method to overestimate the platelet count in samples with 70 fL < MCV ≤ 80 fL, 60 fL < MCV ≤ 70 fL, MCV ≤ 60 fL. Receiver operating characteristic curve analysis showed that a 73.5fL cutoff of MCV was highly sensitive in predicting falsely increased platelet counts.

CONCLUSION

In cases with MCV < 73.5 fL, we strongly suggest that the platelet counts obtained by the impedance method on the Sysmex XN9000 analyzer should be checked and corrected by fluorescence counting.

摘要

目的

尽管小红细胞是一个众所周知的分析陷阱,可能导致血小板计数出现人为增加,但关于小红细胞增多情况下阻抗法血小板计数准确性的信息有限。本研究旨在评估存在小红细胞时阻抗法血小板计数的准确性,并确定认可荧光法血小板计数的最佳平均红细胞体积(MCV)临界值。

方法

在本研究中,将Sysmex XN9000分析仪(Sysmex,日本神户)上通过阻抗法估算的血小板计数与荧光法提供的血小板计数进行比较。评估阻抗法血小板计数的准确性。采用受试者工作特征曲线来评估MCV在预测血小板计数假性增加方面的性能。

结果

在70 fL<MCV≤80 fL、60 fL<MCV≤70 fL、MCV≤60 fL的样本中,阻抗法有高估血小板计数的趋势。受试者工作特征曲线分析表明,MCV临界值为73.5 fL在预测血小板计数假性增加方面具有高度敏感性。

结论

在MCV<73.5 fL的情况下,我们强烈建议对Sysmex XN9000分析仪上通过阻抗法获得的血小板计数进行检查,并通过荧光计数进行校正。