Department of Otolaryngology-Head and Neck Surgery, Faculty of Medicine, Thammasat University, Pathum Thani, Thailand.
Sleep Center of Thammasat (SCENT), Thammasat University Hospital, Pathum Thani, Thailand.
Sleep Breath. 2024 Oct;28(5):1977-1985. doi: 10.1007/s11325-024-03076-3. Epub 2024 Jun 22.
This study aimed to evaluate the accuracy of a Bluetooth position monitor called NaTu sensor and its mobile phone application for detecting sleep position among patients with obstructive sleep apnea (OSA) during polysomnography (PSG).
A cross-sectional study was conducted on adults with suspected of having OSA who underwent PSG. Sleep positions were recorded simultaneously using a video-validated PSG position sensor and the NaTu sensor. The area under the Receiver Operator Characteristic curve (ROC AUC), sensitivity, and specificity values were calculated to evaluate the validity of the NaTu sensor.
Ninety participants (56.7% male) were included, with median age of 40.0 years and body mass index of 29.4 kg/m. The mean AHI was 58.4 ± 31.2 events/hour, categorizing the severity of OSA as mild (5.6%), moderate (18.9%), and severe (75.5%). Sleep positions (supine, lateral right, lateral left) identified by the NaTu sensor closely agreed with the video-validated PSG. The kappa statistic demonstrated almost perfect agreement (k = 0.95, P < 0.001) for overall position recording. The ROC AUC for identifying supine, lateral right, and lateral left positions ranged from 0.974 to 0.981, with sensitivity ranging from 95.1% to 99.1% and specificity from 96.5% to 99.6%.
Our wearable sensor monitoring significantly agrees with video-validated PSG in identifying sleep positions. This device is reliable and accurate for position monitoring and could be an alternative tool for monitoring positions in in-lab PSG, home sleep apnea testing, or tracking positional treatment at home.
Thaiclinicaltrials.org with number TCTR20210701008.
本研究旨在评估一种名为 NaTu 传感器的蓝牙位置监测仪及其移动应用程序在多导睡眠图(PSG)中检测阻塞性睡眠呼吸暂停(OSA)患者睡眠姿势的准确性。
对疑似患有 OSA 并接受 PSG 检查的成年人进行横断面研究。使用经过视频验证的 PSG 位置传感器和 NaTu 传感器同时记录睡眠姿势。计算受试者工作特征曲线(ROC AUC)下的面积、敏感度和特异性值,以评估 NaTu 传感器的有效性。
共纳入 90 名参与者(56.7%为男性),中位年龄 40.0 岁,体重指数为 29.4kg/m。平均 AHI 为 58.4±31.2 次/小时,OSA 严重程度分类为轻度(5.6%)、中度(18.9%)和重度(75.5%)。NaTu 传感器识别的睡眠姿势(仰卧位、右侧卧位、左侧卧位)与视频验证的 PSG 密切一致。kappa 统计量显示整体位置记录具有几乎完美的一致性(k=0.95,P<0.001)。识别仰卧位、右侧卧位和左侧卧位的 ROC AUC 范围为 0.974 至 0.981,敏感度范围为 95.1%至 99.1%,特异性范围为 96.5%至 99.6%。
我们的可穿戴传感器监测在识别睡眠姿势方面与视频验证的 PSG 具有显著一致性。该设备在位置监测方面可靠且准确,可作为实验室 PSG、家庭睡眠呼吸暂停测试或在家中跟踪体位治疗的替代工具。