Suppr超能文献

一种在CZTγ相机上测量带有散射的平面空间分辨率的实用方法。

A practical method for measuring planar spatial resolution with scatter on a CZT gamma camera.

作者信息

Barrau Corinne, Tylski Perrine

机构信息

Unité de Physique Médicale - secteur Médecine Nucléaire, CHU Nîmes, bât ICG, Place du Professeur Robert Debré, Nîmes Cedex 9, 30029, France.

Service de Physique Médicale et Radioprotection, Groupement Hospitalier Est, Hospices Civils de Lyon, Bâtiment Pinel, 52 Bd Pinel, Lyon, Bron, 69677, France.

出版信息

EJNMMI Phys. 2024 Jul 1;11(1):55. doi: 10.1186/s40658-024-00656-w.

Abstract

PURPOSE

This study proposes a practical method for evaluating 2D spatial resolution with scatter on a CZT planar detector gamma camera, which is simpler and faster than the NEMA method. It is used to characterize the influence of distance on spatial resolution FWHM on a CZT camera equipped with a WEHR collimator.

METHODS

The practical method uses linear sources tilted with respect to the detector axes. The spatial resolution full width at half maximum (FWHM) with four tilt angles was compared to the FWHM evaluated using the NEMA NU1-2018 method. Spatial resolution FWHM was also assessed with tilted sources acquired at distances of 0 to 20 cm using a single angle, with and without the post-processing image enhancement proposed by the manufacturer.

RESULTS

Estimated spatial resolution FWHM with tilted sources was close to the spatial resolution FWHM estimated at 7.63 mm by the NEMA method, with deviations ranging from - 5.62 to 4.59% at 10 cm depending on the angle considered. The study of spatial resolution FWHM dependence on distance indicates that, for distances less than 3 cm, the FWHM no longer decreases with distance. The manufacturer's post-processing reduces the FWHM by an average of 15%.

CONCLUSION

The practical method is quicker to implement and gives comparable results to the NEMA reference method for spatial resolution FWHM. Evaluation of spatial resolution with linear sources at short distances from the collimator is limited by the collimator effect and signal digitization. The tilted source method can be used to measure spatial resolution quickly and easily under clinical conditions for CZT planar cameras.

摘要

目的

本研究提出了一种在CZT平面探测器γ相机上评估存在散射情况下二维空间分辨率的实用方法,该方法比NEMA方法更简单、快速。它用于表征距离对配备WEHR准直器的CZT相机上空间分辨率半高宽(FWHM)的影响。

方法

该实用方法使用相对于探测器轴倾斜的线性源。将四个倾斜角度下的空间分辨率半高宽(FWHM)与使用NEMA NU1 - 2018方法评估的FWHM进行比较。还使用单个角度在0至20厘米的距离下采集倾斜源来评估空间分辨率FWHM,有无制造商提出的后处理图像增强。

结果

倾斜源估计的空间分辨率FWHM接近NEMA方法估计的7.63毫米的空间分辨率FWHM,在10厘米处,根据所考虑的角度,偏差范围为 - 5.62%至4.59%。对空间分辨率FWHM与距离的依赖性研究表明,对于小于3厘米的距离,FWHM不再随距离减小。制造商的后处理使FWHM平均降低了15%。

结论

该实用方法实施更快,并且在空间分辨率FWHM方面给出的结果与NEMA参考方法相当。在离准直器短距离处使用线性源评估空间分辨率受准直器效应和信号数字化限制。倾斜源方法可用于在临床条件下快速、轻松地测量CZT平面相机的空间分辨率。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/db7c/11217200/7359c14dd9b4/40658_2024_656_Fig1_HTML.jpg

相似文献

1
A practical method for measuring planar spatial resolution with scatter on a CZT gamma camera.
EJNMMI Phys. 2024 Jul 1;11(1):55. doi: 10.1186/s40658-024-00656-w.
2
Performance characteristics of a new pixelated portable gamma camera.
Med Phys. 2012 Jun;39(6):3435-44. doi: 10.1118/1.4718874.
3
Evaluation of a novel collimator for molecular breast tomosynthesis.
Med Phys. 2017 Nov;44(11):5740-5748. doi: 10.1002/mp.12564. Epub 2017 Oct 13.
4
Analytical and experimental FWHM of a gamma camera: theoretical and practical issues.
PeerJ. 2015 Feb 3;3:e722. doi: 10.7717/peerj.722. eCollection 2015.
6
Characterisation of a hand-held CZT-based gamma camera for Lu imaging.
EJNMMI Phys. 2020 Jul 8;7(1):46. doi: 10.1186/s40658-020-00313-y.
7
Performance characteristics of dedicated molecular breast imaging systems at low doses.
Med Phys. 2016 Jun;43(6):3062-3070. doi: 10.1118/1.4950873.
8
Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator.
J Appl Clin Med Phys. 2021 Feb;22(2):165-177. doi: 10.1002/acm2.13158. Epub 2021 Jan 10.

本文引用的文献

1
Recent Developments in SPECT/CT.
Semin Nucl Med. 2022 May;52(3):276-285. doi: 10.1053/j.semnuclmed.2022.01.004. Epub 2022 Feb 21.
3
Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator.
J Appl Clin Med Phys. 2021 Feb;22(2):165-177. doi: 10.1002/acm2.13158. Epub 2021 Jan 10.
4
Characterisation of a hand-held CZT-based gamma camera for Lu imaging.
EJNMMI Phys. 2020 Jul 8;7(1):46. doi: 10.1186/s40658-020-00313-y.
6
Performance characteristics of a new pixelated portable gamma camera.
Med Phys. 2012 Jun;39(6):3435-44. doi: 10.1118/1.4718874.
7
Fiji: an open-source platform for biological-image analysis.
Nat Methods. 2012 Jun 28;9(7):676-82. doi: 10.1038/nmeth.2019.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验