Suppr超能文献

Heterogeneous enzyme immunoassay with electrochemical detection: competitive and "sandwich"-type immunoassays.

作者信息

Wehmeyer K R, Halsall H B, Heineman W R

出版信息

Clin Chem. 1985 Sep;31(9):1546-9.

PMID:3896571
Abstract

In these competitive and "sandwich"-type heterogeneous enzyme immunoassays, based on liquid chromatography with electrochemical detection, rabbit immunoglobulin G is used as a model compound. Alkaline phosphatase (EC 3.1.3.1), the labeling enzyme, catalyzes conversion of phenyl phosphate to phenol. After separation on an octyldecylsilane column, the enzyme-generated phenol is detected in a thin-layer cell at a carbon-paste working electrode. The detection limit for phenol is 5.0 nmol/L. The electrode response varies linearly with concentration over a range of three orders of magnitude. For the sandwich-type assay procedure the detection limit is 10 ng/L; the linearity ranges over four orders of magnitude. The detection limit of the competitive immunoassay is 5 micrograms/L. The dynamic range spans two orders of magnitude.