J Oral Sci. 2024;66(3):182-188. doi: 10.2334/josnusd.24-0045.
To evaluate the influence of the polymerization distance of monowave and polywave light curing units (LCUs) on the measured irradiance relative to the value reported by the manufacturer in relation to the physical properties of resin-based composites (RBCs).
Four LCUs were used: one monowave and three polywave. The irradiance was measured with a digital radiometer. Depth of cure (DC) and flexural strength (FS) tests were performed according to ISO 4049:2019 at polymerization distances of 0 mm and 5 mm.
The irradiance of all LCUs was higher than that reported by the manufacturer (>25-64%). The irradiance of the four LCUs was reduced when polymerization was performed at between 0 to 5 mm (paired t-test, P < 0.001). The DC at 0 mm was similar in all groups but was significantly decreased at 5 mm distance (ANOVA P < 0.001). FS showed differences among the LCUs at 0 mm (ANOVA P < 0.001) and was affected by the polymerization distance. The elastic modulus was unaffected by the LCU used or the distance (ANOVA P > 0.001).
The LCU must be positioned as near as possible to RBCs during the polymerization process, as increased distance negatively affects the depth of cure and flexural strength.
评估单波和多波光固化器(LCU)的聚合距离对测量辐照度相对于制造商报告值的影响,以研究其与树脂基复合材料(RBC)物理性能的关系。
使用了四种 LCU:一种单波和三种多波。使用数字辐射计测量辐照度。根据 ISO 4049:2019 标准,在聚合距离为 0mm 和 5mm 时进行了固化深度(DC)和弯曲强度(FS)测试。
所有 LCU 的辐照度均高于制造商报告的值(>25-64%)。当在 0 到 5mm 之间进行聚合时,四个 LCU 的辐照度均降低(配对 t 检验,P<0.001)。0mm 处的 DC 在所有组中均相似,但在 5mm 距离处显著降低(ANOVA P<0.001)。FS 在 0mm 处存在 LCUs 之间的差异(ANOVA P<0.001),并受聚合距离的影响。弹性模量不受所用 LCU 或距离的影响(ANOVA P>0.001)。
在聚合过程中,LCU 必须尽可能靠近 RBC 放置,因为距离的增加会对固化深度和弯曲强度产生负面影响。