使用双层光谱 CT 优化单能量图像以减少下颈椎的条纹伪影:一项回顾性研究。

Optimisation of monoenergetic images to reduce banding artifacts in the lower cervical spine using dual-layer spectral computed tomography: a retrospective study.

机构信息

Department of Radiology and Nuclear Medicine, The First Hospital of HeBei Medical University, ShiJiaZhuang 050051, China.

Department of CTMRI, The Third Hospital of HeBei Medical University, ShiJiaZhuang 050051, China.

出版信息

Clin Radiol. 2024 Oct;79(10):e1252-e1259. doi: 10.1016/j.crad.2024.07.001. Epub 2024 Jul 5.

Abstract

AIM

To investigate the application value of dual-layer detector computed tomography (CT) single-energy spectral images for reducing artefacts in the lower cervical spine.

MATERIALS AND METHODS

Sixty-three patients who underwent neck examination using spectral CT between March 2022 and January 2023 were selected as the participants. Conventional mixed-energy images and spectral imaging data at 40-200 KeV were obtained from the spectral CT scans. The standard deviation (SD), signal-to-noise ratio (SNR), |CT| (the difference in CT value between the C6-7 and C3-4 artefact regions), and contrast-to-noise ratio (CNR) of the C6-7 spinal canal were measured. The image quality of the artefact region in the conventional and single images were compared.

RESULTS

The SD value was highest in the single-level 40 KeV images and lowest in the 120 KeV images (P<0.05). The SNR was lowest in the 40 KeV images (2.07 ± 2.74) and highest in the 120 KeV images (6.20 ± 5.26) (P<0.05). The CT was highest in the 40 KeV images and lowest in the 120 KeV images (108.38 vs. 10.63 ± 8.58). The subjective image quality scores were lowest in the 40 KeV images and highest in the 120 KeV images.

CONCLUSION

Our findings suggest that single-energy 120 KeV images may offer benefits such as low SD, elevated SNR, higher CNR, and improved image quality.

摘要

目的

研究双层探测器 CT 单能谱图像在降低下颈椎伪影中的应用价值。

材料和方法

选取 2022 年 3 月至 2023 年 1 月间在我院行颈部 CT 检查的 63 例患者作为研究对象。从能谱 CT 扫描中获得常规混合能量图像和 40-200KeV 的光谱成像数据。测量 C6-7 椎管的标准偏差(SD)、信噪比(SNR)、|CT|(伪影区 C6-7 与 C3-4 的 CT 值差值)和对比噪声比(CNR)。比较常规图像和单能图像伪影区的图像质量。

结果

单能 40KeV 图像的 SD 值最高(P<0.05),120KeV 图像的最低;40KeV 图像的 SNR 最低(2.07±2.74),120KeV 图像的最高(6.20±5.26)(P<0.05);40KeV 图像的 CT 值最高(108.38),120KeV 图像的最低(10.63±8.58);主观图像质量评分 40KeV 图像最低,120KeV 图像最高。

结论

单能 120KeV 图像可能具有低 SD、高 SNR、高 CNR 和改善的图像质量等优势。