To investigate the application value of dual-layer detector computed tomography (CT) single-energy spectral images for reducing artefacts in the lower cervical spine.
Sixty-three patients who underwent neck examination using spectral CT between March 2022 and January 2023 were selected as the participants. Conventional mixed-energy images and spectral imaging data at 40-200 KeV were obtained from the spectral CT scans. The standard deviation (SD), signal-to-noise ratio (SNR), |CT| (the difference in CT value between the C6-7 and C3-4 artefact regions), and contrast-to-noise ratio (CNR) of the C6-7 spinal canal were measured. The image quality of the artefact region in the conventional and single images were compared.
The SD value was highest in the single-level 40 KeV images and lowest in the 120 KeV images (P<0.05). The SNR was lowest in the 40 KeV images (2.07 ± 2.74) and highest in the 120 KeV images (6.20 ± 5.26) (P<0.05). The CT was highest in the 40 KeV images and lowest in the 120 KeV images (108.38 vs. 10.63 ± 8.58). The subjective image quality scores were lowest in the 40 KeV images and highest in the 120 KeV images.
Our findings suggest that single-energy 120 KeV images may offer benefits such as low SD, elevated SNR, higher CNR, and improved image quality.
研究双层探测器 CT 单能谱图像在降低下颈椎伪影中的应用价值。
选取 2022 年 3 月至 2023 年 1 月间在我院行颈部 CT 检查的 63 例患者作为研究对象。从能谱 CT 扫描中获得常规混合能量图像和 40-200KeV 的光谱成像数据。测量 C6-7 椎管的标准偏差(SD)、信噪比(SNR)、|CT|(伪影区 C6-7 与 C3-4 的 CT 值差值)和对比噪声比(CNR)。比较常规图像和单能图像伪影区的图像质量。
单能 40KeV 图像的 SD 值最高(P<0.05),120KeV 图像的最低;40KeV 图像的 SNR 最低(2.07±2.74),120KeV 图像的最高(6.20±5.26)(P<0.05);40KeV 图像的 CT 值最高(108.38),120KeV 图像的最低(10.63±8.58);主观图像质量评分 40KeV 图像最低,120KeV 图像最高。
单能 120KeV 图像可能具有低 SD、高 SNR、高 CNR 和改善的图像质量等优势。