Clinical evaluation of long-term epidural monitoring of intracranial pressure.

作者信息

Gücer G, Viernstein L J, Chubbuck J G, Walker A E

出版信息

Surg Neurol. 1979 Nov;12(5):373-7.

PMID:390751
Abstract

Forty-two epidural sensors for measuring intracranial pressure were implanted in 38 patients over the last four years. The sensor is a passive device consisting of an inductance and a pressure sensitive capacitance resonant at about 50 megahertz. The study involved 17 cases of hydrocephalus, nine of pseudotumor cerebri, and the remainder included cases of cranial trauma, brain tumors, and aneurysms. Four of the sensors failed in patients and were removed. Two of these had to be replaced to continue the pressure monitoring. The intracranial pressure is still being monitored in 22 patients, while 15 have been lost to follow-up for various reasons. Fourteen patients have been followed for about a year, and four for two years. The baseline drift rate in most sensors is between 1 and 2 mm H2O per day, with a few having greater rates of drift. Calibration errors were found in some of the sensors after implantation. These errors were corrected by direct measurement of the cerebral spinal fluid pressure via lumbar puncture. The sensors have proven to be a valuable adjunct in the management and diagnosis of neurosurgical patients by simplifying the intermittent and continuous recording of intracranial pressure. This was so despite the presence of calibration errors and a drift which necessitated an occasional lumbar puncture for making corrections.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验