Department of Ophthalmology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Cardiology School, University of Brescia, Brescia, Italy.
Eye (Lond). 2024 Nov;38(16):3175-3179. doi: 10.1038/s41433-024-03261-4. Epub 2024 Jul 29.
To investigate the presence of silicone oil (SO)-emulsification on the anterior iris surface with anterior segment optical coherence tomography (AS-OCT).
In this single-center cross-sectional study, vitrectomized eyes with SO tamponade that underwent AS-OCT imaging and gonioscopy examination during the postoperative follow-up visits, were reviewed.
45 eyes of 42 consecutive patients were included. In 35.6% of the eyes (n = 16) emulsified SO droplets were detected in the anterior chamber (AC) angle by gonioscopy and in 55.6% (n = 25) on the anterior iris surface by AS-OCT imaging. The presence of SO emulsifications in the AC-angle correlated with the presence of SO emulsifications on the anterior iris surface (OR = 13.4, 95% CI [2.179-82.130]; p = 0.005). The accuracy of the AS-OCT predicting the presence of SO in the AC-angle was 71.0% and the sensitivity was 87.5%. No significant association between SO droplets in the AC and other clinical parameters including endotamponade-duration or type of silicone oil were found. The presence of emulsified SO droplets on the anterior iris surface detected by AS-OCT was significantly correlated to postoperative IOP rise (p = 0.027).
AS-OCT is a suitable method for the detection of SO on the anterior iris surface. SO droplets on the iris surface correlate with elevated postoperative IOP and with the presence of SO in the AC detected by gonioscopy, therefore AS-OCT might be used as a screening method for the detection of SO migration into the AC.
Institutional Review Board of the Hamburg Medical Chamber (Ethik-Kommission der Ärztekammer Hamburg): 2023-300372-WF.
使用眼前节光学相干断层扫描(AS-OCT)检查硅油(SO)乳化在前虹膜表面的存在情况。
在这项单中心横断面研究中,回顾了接受玻璃体切除手术后接受 AS-OCT 成像和房角镜检查的 SO 填塞眼。
共纳入 42 例连续患者的 45 只眼。在 35.6%(n=16)的眼中,房角镜检查发现前房(AC)角中有乳化 SO 液滴,55.6%(n=25)的眼中在虹膜前表面用 AS-OCT 成像发现有 SO 乳化。AC 角中 SO 乳化的存在与虹膜前表面 SO 乳化的存在相关(OR=13.4,95%CI [2.179-82.130];p=0.005)。AS-OCT 预测 AC 中 SO 存在的准确性为 71.0%,敏感性为 87.5%。在 AC 中的 SO 液滴与其他临床参数(包括内填塞持续时间或硅油类型)之间未发现明显相关性。AS-OCT 检测到的虹膜表面乳化 SO 液滴的存在与术后眼压升高显著相关(p=0.027)。
AS-OCT 是检测前虹膜表面 SO 的合适方法。虹膜表面的 SO 液滴与术后眼压升高以及房角镜检查发现的 AC 中的 SO 存在相关,因此 AS-OCT 可能可用作检测 SO 向 AC 迁移的筛查方法。