在动物模型中通过技术评估骨活检质量。

Evaluating bone biopsy quality by technique in an animal model.

作者信息

Ho Corey K, Gimarc David, Carroll Hsieng-Feng, Clay Michael, Schowinsky Jeffrey, Jesse M K, Crawford Amanda M, Marshall Carrie B

机构信息

University of Colorado - Anschutz Medical Campus, Department of Radiology, 12401 E 17th Ave, Aurora, CO 80045, USA.

University of Colorado - Anschutz Medical Campus, Department of Pathology, 12401 E 17th Ave, Aurora, CO 80045, USA.

出版信息

Res Diagn Interv Imaging. 2022 May 21;2:100008. doi: 10.1016/j.redii.2022.100008. eCollection 2022 Jun.

Abstract

RATIONALE AND OBJECTIVES

Powered bone biopsy technique is popular due to its ease of use. However, there is conflicting evidence regarding the diagnostic quality of the samples. The purpose of this study is to evaluate the diagnostic adequacy of different bone biopsy devices and techniques as it relates to the frequency of sample artifacts.

MATERIALS AND METHODS

Bone biopsy was performed on same-day processed lamb femora using the following techniques: manual, pulsed powered and full powered. Ten samples were collected using each method by a single musculoskeletal-trained radiologist and were reviewed by 3 blinded pathologists. Samples were compared across multiple categories: length, bone dust, thermal/crush artifact, cellular morphology, fragmentation, and diagnostic acceptability. Bayesian Multilevel Nonlinear Regression models were performed assessing the association between the techniques across the categories.

RESULTS

Statistical analysis revealed that the manual technique outperformed any powered technique across all categories: decreased thermal/crush artifact ( = 0.014), decreased bone dust (p<0.001), better cellular morphology ( = 0.005), less fragmentation ( < 0.0001) and better diagnostic acceptability ( < 0.0001).

CONCLUSION

Manually obtained bone biopsy samples generally produce a more diagnostic sample as compared to powered techniques in an animal model. Given these results, manual bone biopsy methods should be encouraged after consideration for lesion composition, difficulty of access and the patient's overall condition.

摘要

原理与目的

电动骨活检技术因其使用简便而广受欢迎。然而,关于样本诊断质量的证据存在矛盾。本研究的目的是评估不同骨活检设备和技术的诊断充分性及其与样本伪像频率的关系。

材料与方法

使用以下技术对当天处理的羊股骨进行骨活检:手动、脉冲电动和全电动。由一名接受过肌肉骨骼培训的放射科医生使用每种方法收集10个样本,并由3名不知情的病理学家进行评估。对样本在多个类别上进行比较:长度、骨屑、热/挤压伪像、细胞形态、破碎情况和诊断可接受性。进行贝叶斯多级非线性回归模型分析,评估不同技术在各个类别之间的关联。

结果

统计分析表明,在所有类别中,手动技术均优于任何电动技术:热/挤压伪像减少(P = 0.014),骨屑减少(P<0.001),细胞形态更好(P = 0.005),破碎情况更少(P<0.0001),诊断可接受性更好(P<0.0001)。

结论

在动物模型中,与电动技术相比,手动获取的骨活检样本通常能产生更具诊断价值的样本。鉴于这些结果,在考虑病变组成、取材难度和患者总体状况后,应鼓励采用手动骨活检方法。