BACKGROUND

Most acute coronary syndromes occur due to coronary thrombosis caused by plaque rupture (PR) and plaque erosion (PE). Precise differentiation between PR and PE is challenging for intravascular imaging. This study is the first to determine the diagnostic performance of the novel 60 MHz high-definition intravascular ultrasound (HD-IVUS) for differentiating atherosclerotic plaque morphology influenced by local hemodynamic flow in rabbits. This study evaluated the diagnostic performance of 60 MHz HD-IVUS in identifying thrombosis in rabbits.

METHODS

We established 60 rabbit models of atherosclerosis with left common carotid artery (LCCA) stenosis and 30 -induced LCCA thrombosis. Intravascular imaging was assessed with 60 MHz HD-IVUS and fourier-domain optical coherence tomography (FD-OCT). The present study investigated the diagnostic accuracy of 60 MHz HD-IVUS for PR and PE, as well as thrombosis, using OCT-diagnosis as a standard reference.

RESULTS

60 MHz HD-IVUS for identifying atherosclerotic plaque morphology using plaque cavity and minor intimal irregularities showed high sensitivity and specificity; 92.0 and 90.0% for identifying OCT-defined PR, and 80.0 and 70.0% for OCT-defined PE, respectively. In a rabbit thrombus model, 60 MHz HD-IVUS showed high sensitivity (88.0%) and specificity (80.0%) in identifying OCT-defined thrombosis.

CONCLUSIONS

60 MHz HD-IVUS can accurately identify PR and thrombosis. Further studies should confirm the clinical value of this novel technique in PE diagnosis.

摘要

背景

大多数急性冠状动脉综合征是由斑块破裂(PR)和斑块侵蚀(PE)导致的冠状动脉血栓形成引起的。对于血管内成像而言,精确区分PR和PE具有挑战性。本研究首次确定新型60MHz高清血管内超声(HD-IVUS)在区分受局部血流动力学影响的兔动脉粥样硬化斑块形态方面的诊断性能。本研究评估了60MHz HD-IVUS在识别兔血栓形成方面的诊断性能。

方法

我们建立了60只左颈总动脉(LCCA)狭窄的动脉粥样硬化兔模型和30只诱导的LCCA血栓形成模型。使用60MHz HD-IVUS和傅里叶域光学相干断层扫描(FD-OCT)进行血管内成像评估。本研究以OCT诊断为标准参考,研究了60MHz HD-IVUS对PR、PE以及血栓形成的诊断准确性。

结果

60MHz HD-IVUS利用斑块腔和内膜微小不规则性识别动脉粥样硬化斑块形态时显示出高敏感性和特异性;识别OCT定义的PR时分别为92.0%和90.0%,识别OCT定义的PE时分别为80.0%和70.0%。在兔血栓模型中,60MHz HD-IVUS在识别OCT定义的血栓形成方面显示出高敏感性(88.0%)和特异性(80.0%)。

结论

60MHz HD-IVUS能够准确识别PR和血栓形成。进一步的研究应证实这项新技术在PE诊断中的临床价值。