Fischer M, Alexander K
Stroke. 1985 Nov-Dec;16(6):973-6. doi: 10.1161/01.str.16.6.973.
Doppler sonography of 29 extracranial carotid arteries was performed twice within three days or less. Angiography revealed stenoses with from 10 to 95% diameter reduction in 25 internal carotid arteries whereas four vessels were found to be normal. The systolic peak frequency of the internal carotid artery read from the Doppler spectrum (n = 29) could be reproduced very well. This was shown by a linear regression nearest the line of identical values with a coefficient of correlation r = 0.97 (p less than 0.001). Other values derived from the spectral analysis of the Doppler shift signal were not so well reproduced including the peak frequency ratio (systolic peak frequency of the internal carotid artery/systolic peak of the common carotid artery) (n = 22; r = 0.81; p less than 0.001). The mean frequencies read from the zero-crossing detector recordings (n = 20) could not be reproduced as demonstrated by a linear regression far away from the line of identical measurements with a coefficient of correlation r = 0.43 (p less than 0.05).
对29条颅外颈动脉在三天或更短时间内进行了两次多普勒超声检查。血管造影显示,25条颈内动脉存在直径缩小10%至95%的狭窄,而另外4条血管正常。从多普勒频谱读取的颈内动脉收缩期峰值频率(n = 29)能够很好地重现。这通过最接近恒值线的线性回归得以体现,相关系数r = 0.97(p < 0.001)。从多普勒频移信号频谱分析得出的其他值重现效果欠佳,包括峰值频率比(颈内动脉收缩期峰值频率/颈总动脉收缩期峰值)(n = 22;r = 0.81;p < 0.001)。从零交叉检测器记录读取的平均频率(n = 20)无法重现,线性回归显示其远离恒值测量线,相关系数r = 0.43(p < 0.05)。