含金属植入物的第二代双层光谱探测器 CT 图像的碘浓度、HU 准确性和金属伪影评估:一项体模研究。

Iodine concentration, HU accuracy, and metal artifacts evaluation on second-generation dual-layer spectral detector CT images with metal implants: a phantom study.

机构信息

Department of Clinical Engineering, Nanfang Hospital, Southern Medical University, Guangzhou, PR China.

School of Biomedical Engineering, Southern Medical University, Guangzhou, PR China.

出版信息

Acta Radiol. 2024 Sep;65(9):1030-1038. doi: 10.1177/02841851241269853. Epub 2024 Aug 14.

Abstract

BACKGROUND

Metal implants may affect the image quality, iodine concentration (IC), and CT Hounsfield unit (HU) quantification accuracy.

PURPOSE

To investigate the quantitative accuracy of IC and HU from dual-layer spectral detector (DLCT) in the presence of metal artifacts.

MATERIAL AND METHODS

An experimental cylindrical phantom containing eight iodine inserts and two metal inserts was designed. The phantom underwent scanning at three radiation dose levels and two tube voltage settings. A set of conventional images (CIs), virtual monoenergetic images (VMIs), and iodine concentration maps (ICMs) were generated and measured for all the eight iodine inserts. Quantitative indicators of mean absolute percentage error (MAPE), artifact index (AI), contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and standard deviation (SD) on CIs and VMIs were calculated for IC and HU. Subjective score evaluation was also conducted.

RESULTS

The MAPE values of all regions of interest across different scanning configurations were all <5%. Almost all APE values were <5%, indicating that metal artifacts had little impact on IC measurements. When the tube voltage was fixed, the SD value of attenuation decreased with the increase of the tube current; this is also true when the tube current was fixed. The middle energy reconstructions seemed to give a good balance between reducing artifacts and improving contrast.

CONCLUSION

VMIs from DLCT can reduce metal artifacts, the accuracy of IC quantification is not sensitive to imaging parameters. In summary, metal implants exhibit minimal impact on image quality and IC quantification accuracy in reconstructed images from DLCT.

摘要

背景

金属植入物可能会影响图像质量、碘浓度 (IC) 和 CT 亨氏单位 (HU) 的定量准确性。

目的

研究双层光谱探测器 (DLCT) 在存在金属伪影时的 IC 和 HU 定量准确性。

材料和方法

设计了一个包含八个碘插塞和两个金属插塞的圆柱形实验体模。体模在三个辐射剂量水平和两个管电压设置下进行扫描。对所有八个碘插塞生成并测量了一组常规图像 (CIs)、虚拟单能量图像 (VMIs) 和碘浓度图 (ICMs)。计算了 CIs 和 VMIs 上 IC 和 HU 的平均绝对百分比误差 (MAPE)、伪影指数 (AI)、对比噪声比 (CNR)、信噪比 (SNR) 和标准偏差 (SD) 的定量指标。还进行了主观评分评估。

结果

不同扫描配置下所有感兴趣区域的 MAPE 值均<5%。几乎所有 APE 值均<5%,表明金属伪影对 IC 测量的影响很小。当管电压固定时,衰减的 SD 值随管电流的增加而降低;当管电流固定时也是如此。中间能量重建似乎在减少伪影和提高对比度之间取得了良好的平衡。

结论

DLCT 的 VMIs 可以减少金属伪影,IC 定量的准确性对成像参数不敏感。总之,金属植入物对 DLCT 重建图像的图像质量和 IC 定量准确性的影响很小。