Suppr超能文献

用于创伤性损伤手腕血管重建的微血管吻合耦合器的效率与安全性

Efficiency and safety of microvascular anastomotic coupler for wrist revascularization in traumatic injuries.

作者信息

Mai Tuong Trong, Nguyen Ly Thi Thanh, Nguyen Phi Duong

机构信息

Microsurgery and Limb Reconstruction Department, Hospital for Traumatology and Orthopaedics, Ho Chi Minh City, Vietnam.

Orthopaedic - Burn - Plastic Surgery Department, City Children's Hospital, Ho Chi Minh City, Vietnam.

出版信息

JPRAS Open. 2024 Jul 5;41:252-259. doi: 10.1016/j.jpra.2024.06.017. eCollection 2024 Sep.

Abstract

INTRODUCTION

Effective vascular anastomosis is crucial for successful hand surgery, particularly in traumatic injuries where rapid restoration of blood flow is essential. The Synovis microvascular anastomotic coupler system presents a novel approach by potentially offering faster and more reliable outcomes than traditional suturing methods. This study was conducted to assess the effectiveness and safety of the Synovis coupler for microvascular anastomoses in the wrist, evaluating anastomosis time, patency rates, and complication rates.

METHODS

This retrospective study examined 25 microvascular anastomoses (22 arteries and 3 veins) in 24 patients at a major trauma center. The primary outcomes measured were anastomosis time and vascular patency, which were assessed immediately post-operation and at subsequent follow-ups.

RESULTS

The average time required for anastomosis using the coupler was 7.3 min. Immediate post-operative vascular patency was 100%, with a long-term patency rate of 88%. Complications included arterial narrowing at the coupler site in 3 arteries, and complete occlusion in 3 arteries, accounting for a 12% long-term complication rate.

CONCLUSION

Although the Synovis coupler demonstrated a significant reduction in anastomosis time and high patency rates, our data suggest that its effectiveness is more pronounced in venous anastomoses. The higher complication rates in arterial anastomoses warrant cautious use in these cases. Further research, including prospective randomized controlled trials, is needed to validate these findings and optimize patient selection criteria for using couplers in microvascular surgery.

摘要

引言

有效的血管吻合对于手部手术的成功至关重要,特别是在创伤性损伤中,快速恢复血流至关重要。Synovis微血管吻合耦合器系统提供了一种新方法,可能比传统缝合方法产生更快、更可靠的结果。本研究旨在评估Synovis耦合器在腕部微血管吻合中的有效性和安全性,评估吻合时间、通畅率和并发症发生率。

方法

这项回顾性研究检查了一家主要创伤中心24例患者的25例微血管吻合(22条动脉和3条静脉)。测量的主要结果是吻合时间和血管通畅情况,在术后立即及随后的随访中进行评估。

结果

使用耦合器进行吻合的平均时间为7.3分钟。术后立即血管通畅率为100%,长期通畅率为88%。并发症包括3条动脉在耦合器部位出现动脉狭窄,3条动脉完全闭塞,长期并发症发生率为12%。

结论

尽管Synovis耦合器显示吻合时间显著缩短且通畅率高,但我们的数据表明其在静脉吻合中的有效性更为明显。动脉吻合中较高的并发症发生率提示在这些病例中应谨慎使用。需要进一步的研究,包括前瞻性随机对照试验,以验证这些发现并优化在微血管手术中使用耦合器的患者选择标准。