Indican interference with six commercial procedures for measuring total bilirubin.

作者信息

Poon R, Hinberg I H

出版信息

Clin Chem. 1985 Jan;31(1):92-4.

PMID:3917384
Abstract

We have studied the effect of indican on six commercial procedures for the measurement of total bilirubin in serum. Total bilirubin measured by the Bilirubin A-Gent (Abbott) 2,4-dichlorophenyl diazonium procedure increased by 50 mg/L for each 1 mmol/L of added indican. Similarly, total bilirubin measured by the Bilirubin C-System (Boehringer Mannheim) 2,5-dichlorophenyl diazonium procedure increased by 33 mg/L per mmol/L of indican. Indican also interfered with the Micro Bilirubin Reagent Set (Harleco) Malloy-Evelyn procedure, but to a much lesser extent. The Jendrassik Bilirubin Reagent System (American Monitor) and a modified Jendrassik-Grof procedure (Hoffmann-LaRoche) adapted to the Cobas Bio analyzer were unaffected by the presence of indican. The amount of interference with the 2,5-dichlorophenyl diazonium procedure increased significantly with color development time and was twice the initial amount after 30 min. Concentrations of indican as high as 0.38 mmol/L have been found in sera of patients with renal failure, which would increase total bilirubin values measured by the first two procedures above by 19 and 12 mg/L, respectively. Users of these procedures should therefore be suspicious of unexpectedly high bilirubin values obtained with sera from patients with chronic renal disease.

摘要

相似文献

1
Indican interference with six commercial procedures for measuring total bilirubin.
Clin Chem. 1985 Jan;31(1):92-4.
2
A 2,4-dichlorophenyl diazonium-based method for total bilirubin without interference from indican in uremic sera.一种用于测定总胆红素的基于2,4-二氯苯基重氮盐的方法,该方法不受尿毒症血清中吲哚酚的干扰。
Clin Chem. 1985 Jul;31(7):1229-31.
3
Indican interference in bilirubin assays: a classical solution still applies.靛基质对胆红素检测的干扰:经典解决方案仍然适用。
Clin Chem. 2007 Feb;53(2):359-61. doi: 10.1373/clinchem.2006.083154.
4
Diazo-based assay for total bilirubin in a coated thin film evaluated.
Clin Chem. 1984 Aug;30(8):1310-3.
5
Automated Jendrassik-Grof method for measurement of bilirubin in serum with the Greiner Selective Analyzer (GSA II D), and comparison with the method involving diazotized 2,4-dichloroaniline.使用格赖纳选择性分析仪(GSA II D)自动测定血清胆红素的Jendrassik-Grof方法,以及与重氮化2,4-二氯苯胺方法的比较。
Clin Chem. 1981 May;27(5):748-52.
6
A modified Malloy-Evelyn procedure for total bilirubin in microsamples.一种用于微量样本总胆红素检测的改良马洛伊-伊夫林法。
Am J Med Technol. 1977 Oct;43(10):1004-7.
7
Inaccurate values for direct bilirubin with some commonly used direct bilirubin procedures.
Clin Chem. 1985 Sep;31(9):1560-3.
8
Naproxen metabolites interfere with certain bilirubin assays: elimination of interference by using a Roche bilirubin assay on the Hitachi 917 analyzer.萘普生代谢物会干扰某些胆红素检测:通过在日立 917 分析仪上使用罗氏胆红素检测消除干扰。
Am J Clin Pathol. 2010 Jun;133(6):878-83. doi: 10.1309/AJCPN6MWATQ3SZTC.
9
A modification of the acid diazo coupling method (Malloy-Evelyn) for the determination of serum total bilirubin.
Scand J Clin Lab Invest. 1997 May;57(3):275-9. doi: 10.3109/00365519709060037.
10
One-step protocol for assays of total and direct bilirubin with stable combined reagents.
Clin Chem. 1989 Jan;35(1):173-6.