Indican interference with six commercial procedures for measuring total bilirubin.

作者信息

Poon R, Hinberg I H

出版信息

Clin Chem. 1985 Jan;31(1):92-4.

PMID:3917384
Abstract

We have studied the effect of indican on six commercial procedures for the measurement of total bilirubin in serum. Total bilirubin measured by the Bilirubin A-Gent (Abbott) 2,4-dichlorophenyl diazonium procedure increased by 50 mg/L for each 1 mmol/L of added indican. Similarly, total bilirubin measured by the Bilirubin C-System (Boehringer Mannheim) 2,5-dichlorophenyl diazonium procedure increased by 33 mg/L per mmol/L of indican. Indican also interfered with the Micro Bilirubin Reagent Set (Harleco) Malloy-Evelyn procedure, but to a much lesser extent. The Jendrassik Bilirubin Reagent System (American Monitor) and a modified Jendrassik-Grof procedure (Hoffmann-LaRoche) adapted to the Cobas Bio analyzer were unaffected by the presence of indican. The amount of interference with the 2,5-dichlorophenyl diazonium procedure increased significantly with color development time and was twice the initial amount after 30 min. Concentrations of indican as high as 0.38 mmol/L have been found in sera of patients with renal failure, which would increase total bilirubin values measured by the first two procedures above by 19 and 12 mg/L, respectively. Users of these procedures should therefore be suspicious of unexpectedly high bilirubin values obtained with sera from patients with chronic renal disease.

摘要

相似文献

9
A modification of the acid diazo coupling method (Malloy-Evelyn) for the determination of serum total bilirubin.
Scand J Clin Lab Invest. 1997 May;57(3):275-9. doi: 10.3109/00365519709060037.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验