Suppr超能文献

用于评估植入心脏器械患者心肌瘢痕的宽带射频脉冲序列

Wideband radiofrequency pulse sequence for evaluation of myocardial scar in patients with cardiac implantable devices.

作者信息

Shah Neil D, Krishnam Mayil, Ambale Venkatesh Bharat, Khan Fouzia, Smith Michele, Jones Darwin R, Koon Patrick, Mao Xianglun, Janich Martin A, Brau Anja C S, Salerno Michael, Dash Rajesh, Chan Frandics, Yang Phillip C

机构信息

Department of Medicine, Medical College of Wisconsin, Milwaukee, WI, United States.

Department of Radiology, Stanford University, Stanford, CA, United States.

出版信息

Front Radiol. 2024 Aug 7;4:1327406. doi: 10.3389/fradi.2024.1327406. eCollection 2024.

Abstract

BACKGROUND

Cardiac magnetic resonance is a useful clinical tool to identify late gadolinium enhancement in heart failure patients with implantable electronic devices. Identification of LGE in patients with CIED is limited by artifact, which can be improved with a wide band radiofrequency pulse sequence.

OBJECTIVE

The authors hypothesize that image quality of LGE images produced using wide-band pulse sequence in patients with devices is comparable to image quality produced using standard LGE sequences in patients without devices.

METHODS

Two independent readers reviewed LGE images of 16 patients with CIED and 7 patients without intracardiac devices to assess for image quality, device-related artifact, and presence of LGE using the American Society of Echocardiography/American Heart Association 17 segment model of the heart on a 4-point Likert scale. The mean and standard deviation for image quality and artifact rating were determined. Inter-observer reliability was determined by calculating Cohen's kappa coefficient. Statistical significance was determined by -test as a {less than or equal to} 0.05 with a 95% confidence interval.

RESULTS

All patients underwent CMR without any adverse events. Overall IQ of WB LGE images was significantly better in patients with devices compared to standard LGE in patients without devices ( = 0.001) with reduction in overall artifact rating ( = 0.05).

CONCLUSION

Our study suggests wide-band pulse sequence for LGE can be applied safely to heart failure patients with devices in detection of LV myocardial scar while maintaining image quality, reducing artifact, and following routine imaging protocol after intravenous gadolinium contrast administration.

摘要

背景

心脏磁共振成像(CMR)是一种有用的临床工具,可用于识别植入式电子设备的心力衰竭患者的延迟钆增强(LGE)。心脏植入电子设备(CIED)患者的LGE识别受伪影限制,而宽带射频脉冲序列可改善这种情况。

目的

作者假设,在有设备的患者中使用宽带脉冲序列生成的LGE图像的图像质量与在无设备的患者中使用标准LGE序列生成的图像质量相当。

方法

两名独立的阅片者对16例CIED患者和7例无心脏内装置的患者的LGE图像进行评估,使用美国超声心动图学会/美国心脏协会的17节段心脏模型,以4分李克特量表评估图像质量、与设备相关的伪影以及LGE的存在情况。确定图像质量和伪影评级的平均值和标准差。通过计算科恩kappa系数确定观察者间的可靠性。通过t检验确定统计学显著性,α≤0.05,置信区间为95%。

结果

所有患者均接受CMR检查,未发生任何不良事件。与无设备患者的标准LGE相比,有设备患者的宽带LGE图像的总体图像质量明显更好(P = 0.001),总体伪影评级降低(P = 0.05)。

结论

我们的研究表明,用于LGE的宽带脉冲序列可安全应用于有设备的心力衰竭患者,以检测左心室心肌瘢痕,同时保持图像质量、减少伪影,并在静脉注射钆造影剂后遵循常规成像方案。