Suppr超能文献

[热释光剂量测定法对高能电子响应的能量依赖性]

[Energy dependence of the response of thermoluminescent dosimetry to high-energy electrons].

作者信息

Lübbert K, Rahim H

出版信息

Strahlentherapie. 1985 Feb;161(2):109-12.

PMID:3919469
Abstract

Sufficiently reproducible measuring results can be achieved even for simple TLD units by a control procedure which is easy to apply. It is demonstrated by means of the proposed control procedure that the dependence on energy of the TLD-100 rods is negligible in case of a measuring uncertainty of +/- 3%. This is confirmed by depth dose measurements in comparison with ionisation measurements and semiconductor measurements. TLD-100 rods are thus suitable for the measurement of the dose distribution in a typical solid phantom, such as the humanoid phantom of Alderson Research Laboratories, Long Island, N.Y. The results of these measurements, especially with regard to the verification of the irradiation planning, will be presented in a later report.

摘要

即使对于简单的热释光剂量计(TLD)单元,通过易于应用的控制程序也能获得足够可重复的测量结果。通过所提出的控制程序表明,在测量不确定度为±3%的情况下,TLD - 100棒对能量的依赖性可忽略不计。与电离测量和半导体测量相比,深度剂量测量证实了这一点。因此,TLD - 100棒适用于测量典型固体体模中的剂量分布,例如纽约长岛奥尔德森研究实验室的人体模型。这些测量结果,特别是关于照射计划验证的结果,将在后续报告中呈现。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验