构建方向对采用数字光处理制造的义齿牙齿耐磨性和硬度的影响:一项体外研究。

Effect of build orientation on the wear resistance and hardness of denture teeth fabricated using digital light processing: An in vitro study.

作者信息

Sone Khin Pyae, Takahashi Hidekazu, Iwaki Maiko, Namano Sahaprom, Komagamine Yuriko, Minakuchi Shunsuke, Kanazawa Manabu

机构信息

Gerodontology and Oral Rehabilitation, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, Tokyo, Japan.

School of Oral Health Engineering, Faculty of Dentistry, Tokyo Medical and Dental University, Tokyo, Japan.

出版信息

J Prosthodont Res. 2025 Apr 14;69(2):267-274. doi: 10.2186/jpr.JPR_D_24_00111. Epub 2024 Aug 27.

Abstract

PURPOSE

This in vitro study investigated the effect of build orientation on the wear resistance and hardness of denture teeth fabricated using digital light processing (DLP) compared to other denture tooth materials.

METHODS

Disc-shaped specimens were prepared using denture tooth monomers and DLP devices in three build orientations: 0°, 45°, and 90°. Specimens of the same shape were fabricated using denture tooth materials for subtractive manufacturing, commercially available polymethylmethacrylate (PMMA) resin, and composite resin. The wear resistance was evaluated as the wear volume loss after 50,000 wear cycles using a ball-on-disc wear device in water for two-body wear and poppy seed slurry for three-body wear. The Vickers hardness values of the materials were measured. Two-way and one-way analyses of variance were performed for wear resistance and hardness, respectively, followed by Tukey's honest significance test.

RESULTS

The interaction between the denture tooth resins and maximum wear volume was significant (P < 0.01). The 0° build orientation exhibited the lowest wear volume in the three-body wear test and the highest hardness among the three build orientations. The 0° DLP-fabricated specimens demonstrated significantly less wear volume than that of the PMMA specimens and a wear volume comparable to that of the milled specimens. However, the 0° DLP-fabricated specimens showed significantly lower hardness than that of the milled and PMMA specimens. The composite resin specimens exhibited the highest wear resistance and hardness.

CONCLUSIONS

A 0° build orientation is recommended for DLP-fabricated denture teeth compared to 45° and 90° orientations to achieve greater wear resistance and hardness.

摘要

目的

本体外研究调查了与其他义齿材料相比,成型方向对采用数字光处理(DLP)制作的义齿的耐磨性和硬度的影响。

方法

使用义齿单体和DLP设备,在三个成型方向(0°、45°和90°)制备圆盘形试样。使用用于减材制造的义齿材料、市售聚甲基丙烯酸甲酯(PMMA)树脂和复合树脂制作相同形状的试样。使用销盘磨损装置,在水中进行两体磨损试验,磨损50,000个循环后,以及在罂粟籽浆液中进行三体磨损试验后,评估耐磨性,以磨损体积损失表示。测量材料的维氏硬度值。分别对耐磨性和硬度进行双向和单向方差分析,随后进行Tukey诚实显著性检验。

结果

义齿树脂与最大磨损体积之间的相互作用显著(P < 0.01)。在三体磨损试验中,0°成型方向的磨损体积最低,且在三个成型方向中硬度最高。0°方向DLP制作的试样磨损体积明显小于PMMA试样,与铣削试样的磨损体积相当。然而,0°方向DLP制作的试样硬度明显低于铣削试样和PMMA试样。复合树脂试样表现出最高的耐磨性和硬度。

结论

与45°和90°方向相比,DLP制作义齿时,建议采用0°成型方向,以获得更高的耐磨性和硬度。