Gąsior Jakub S, Młyńczak Marcel, Rosoł Maciej, Wieniawski Piotr, Pietrzak Radosław, Werner Bożena
Department of Pediatric Cardiology and General Pediatrics, Medical University of Warsaw, 02-091 Warsaw, Poland.
Institute of Metrology and Biomedical Engineering, Warsaw University of Technology, 02-525 Warsaw, Poland.
J Clin Med. 2024 Aug 8;13(16):4654. doi: 10.3390/jcm13164654.
: Wearable technologies have been developed to measure physiological parameters conveniently. To consider the new measurement device valid, the crucial point is to assess its reliability with the gold standard. The study aimed to assess the validity of the Pneumonitor (PM, fs = 250 Hz) for acquisition of 5 min RR intervals (RRi) for analysis of heart rate asymmetry (HRA) in relation to the electrocardiography (ECG, fs = 1000 Hz) in a group of 19 pediatric cardiac patients. Association between HRA and respiratory rate (RespRate) was verified. : The validation comprised Bland-Altman analysis, intraclass correlation coefficient, and Student's -test. : Sufficient agreement between 10 from 16 HRA parameters was observed. Different HRA parameters values calculated based on RRi from both devices were related to different results of correlation analysis between two parameters and RespRate. : The PM might be considered valid for recording RRi, which are then processed to calculate selected HRA parameters in a group of pediatric cardiac patients in rest condition. However, RRi recorded using devices with fs < 250 Hz may be not adequate for reliable HRA analysis.
可穿戴技术已被开发用于方便地测量生理参数。要使新的测量设备有效,关键在于用金标准评估其可靠性。该研究旨在评估在一组19名儿科心脏病患者中,Pneumonitor(PM,采样频率fs = 250 Hz)采集5分钟心率间期(RRi)以分析心率变异性(HRA)相对于心电图(ECG,采样频率fs = 1000 Hz)的有效性。验证了HRA与呼吸频率(RespRate)之间的关联。验证包括Bland-Altman分析、组内相关系数和学生t检验。观察到16个HRA参数中的10个之间有足够的一致性。基于两种设备的RRi计算出的不同HRA参数值与两个参数和RespRate之间的不同相关性分析结果相关。对于记录RRi,PM可能被认为是有效的,然后在一组处于静息状态的儿科心脏病患者中对RRi进行处理以计算选定的HRA参数。然而,使用采样频率fs < 250 Hz的设备记录的RRi可能不足以进行可靠的HRA分析。