Kardiol Pol. 2023;81(5):491-500. doi: 10.33963/KP.a2023.0070. Epub 2023 Mar 16.
Breathing pattern alterations change the variability and spectral content of the RR intervals (RRi) on electrocardiogram (ECG). However, there is no method to record and control participants' breathing without influencing its natural rate and depth in heart rate variability (HRV) studies.
This study aimed to assess the validity of the Pneumonitor for acquisition of short-term (5 minutes) RRi in comparison to the reference ECG method for analysis of heart rate (HR) and HRV parameters in the group of pediatric patients with cardiac disease.
Nineteen patients of both sexes participated in the study. An ECG and Pneumonitor were used to record RRi in 5-minute static rest conditions, the latter also to measure the relative tidal volume and respiratory rate. The validation comprised Student's t-test, Bland-Altman analysis, intraclass correlation coefficient, and Lin's concordance correlation. The possible impact of respiratory activity on the agreement between ECG and the Pneumonitor was also assessed.
An acceptable agreement for the number of RRi, mean RR, hazard ratio (HR), and HRV measures calculated based on RRi acquired using the ECG and Pneumonitor was presented. There was no association between the breathing pattern and RRi agreement between devices.
The Pneumonitor might be considered appropriate for cardiorespiratory studies in the group of pediatric cardiac patients in rest condition.
呼吸模式的改变会改变心电图(ECG)中 RR 间期(RRi)的可变性和频谱内容。然而,在心率变异性(HRV)研究中,没有一种方法可以在不影响其自然频率和深度的情况下记录和控制参与者的呼吸。
本研究旨在评估 Pneumonitor 在获取短期(5 分钟)RRi 方面的有效性,与参考 ECG 方法相比,该方法用于分析心脏病患儿组的心率(HR)和 HRV 参数。
19 名男女患者参加了这项研究。使用 ECG 和 Pneumonitor 在 5 分钟的静态休息条件下记录 RRi,后者还测量相对潮气量和呼吸频率。验证包括学生 t 检验、Bland-Altman 分析、组内相关系数和林氏一致性相关。还评估了呼吸活动对 ECG 和 Pneumonitor 之间 RRi 一致性的可能影响。
RRi 数量、平均 RR、危险比(HR)和基于 ECG 和 Pneumonitor 获得的 RRi 计算的 HRV 测量值的可接受一致性。设备之间 RRi 一致性与呼吸模式之间没有关联。
在休息状态下的儿科心脏病患者群体中,Pneumonitor 可用于心肺呼吸研究。