Rogulski Mateusz, Pałac Małgorzata, Wolny Tomasz, Linek Paweł
Musculoskeletal Diagnostic and Physiotherapy-Research Team, The Jerzy Kukuczka Academy of Physical Education, 40-065 Katowice, Poland.
Musculoskeletal Elastography and Ultrasonography Laboratory, Institute of Physiotherapy and Health Sciences, The Jerzy Kukuczka Academy of Physical Education, 40-065 Katowice, Poland.
Diagnostics (Basel). 2024 Aug 14;14(16):1771. doi: 10.3390/diagnostics14161771.
There is no validated method of assessing masseter muscle thickness (MMT) by ultrasound imaging (US). However, this is important to ensure study and measurement quality of MMT by US in future studies, as MMT differs depending on the examined area. Thus, this study's aim was to present a new standardized method for assessing the MMT by US and to evaluate the reliability, consistency, and accuracy of its measurements. We also compared the results of MMT measurements obtained by US and computer tomography (CT). The study included nine healthy adults. The US and CT scans were collected in a supine rest position with the mandible in relaxed position. US measurements were determined according to a new standardized protocol (with precise probe location). The MMT measured by CT and US over a seven-day interval showed excellent intra-rater reliability. The mean MMT measured by CT was 12.1 mm (1.74) on the right side and 11.9 mm (1.61) on the left side. The mean MMT measured by US was 12.7 mm (2.00) on the right side and 11.5 mm (1.37) on the left side. The mean percent error in MMT measurement between CT and US was below 6%. A strong linear relationship was found between the CT and US measurements of the MMT on both body sides ( < 0.001, r ≥ 0.93). The proposed method of MMT measurement using US demonstrated excellent reliability, yielding results similar to those obtained from CT images. We recommend the use of this standardization protocol in further studies where precise assessment of MMT by US is expected.
目前尚无通过超声成像(US)评估咬肌厚度(MMT)的有效方法。然而,由于MMT会因检查区域的不同而有所差异,因此在未来的研究中,确保通过US对MMT进行研究和测量的质量非常重要。因此,本研究的目的是提出一种通过US评估MMT的新标准化方法,并评估其测量的可靠性、一致性和准确性。我们还比较了通过US和计算机断层扫描(CT)获得的MMT测量结果。该研究纳入了9名健康成年人。在仰卧休息位且下颌处于放松状态下收集US和CT扫描图像。US测量是根据一种新的标准化方案(精确的探头位置)进行的。在七天的时间间隔内,通过CT和US测量的MMT显示出极好的评分者内可靠性。通过CT测量的右侧MMT平均为12.1 mm(1.74),左侧为11.9 mm(1.61)。通过US测量的右侧MMT平均为12.7 mm(2.00),左侧为11.5 mm(1.37)。CT和US之间MMT测量的平均百分比误差低于6%。在身体两侧,CT和US对MMT的测量之间发现了很强的线性关系(<0.001,r≥0.93)。所提出的使用US测量MMT的方法显示出极好的可靠性,其结果与从CT图像获得的结果相似。我们建议在预期通过US精确评估MMT的进一步研究中使用该标准化方案。