A diffusion-prepared reduced FOV sequence for prostate MRI near metallic implants.

机构信息

Radiology, Stanford University, Stanford, California, USA.

Bioengineering, Stanford University, Stanford, California, USA.

出版信息

Magn Reson Med. 2025 Jan;93(1):261-275. doi: 10.1002/mrm.30280. Epub 2024 Sep 2.

Abstract

PURPOSE

To enable diffusion weighted imaging in prostate patients with metallic total hip replacements in clinically feasible scan times for prostate cancer screening, and avoid distortion and dropout artifacts present in the conventionally used Echo Planar Imaging (EPI).

METHODS

A reduced field of view (FOV) diffusion-prepared sequence that is robust to the B inhomogeneities produced by total hip replacements was achieved using high radiofrequency (RF) bandwidth pulses and manipulation for stimulated echo pathways. The reduced FOV along the A/P direction was obtained using slice-select gradient reversal, and the prepared magnetization was imaged with a three-dimensional RF-spoiled gradient echo readout. The sequence was validated in phantom experiments, in vivo in healthy volunteers with and without total hip replacements, and in vivo in patients undergoing a standard MRI prostate exam.

RESULTS

The proposed sequence is robust to shading and distortion artifacts that are encountered by standard diffusion-weighted EPI in the presence of moderate off-resonance. Apparent diffusion coefficient estimates obtained by the proposed sequence were comparable to those obtained with diffusion-weighted EPI.

CONCLUSION

Acquisition of distortionless diffusion weighted images of the prostate is feasible in patients with total hip replacements on conventional, whole-body 3T MRI, using a b-value of 800 and nominal resolution of 1.7 1.7 4 mm in scan times of 6 min.

摘要

目的

为了在临床上可行的前列腺癌筛查扫描时间内,为患有金属全髋关节置换的前列腺患者实现弥散加权成像,并且避免传统使用的平面回波成像(EPI)中存在的失真和失超伪影。

方法

使用高射频(RF)带宽脉冲和对受激回波途径的操作,实现了对全髋关节置换产生的 B 不均匀性具有鲁棒性的缩小视野(FOV)弥散准备序列。沿着 A/P 方向的缩小 FOV 通过切片选择梯度反转来实现,准备的磁化通过三维 RF 扰相梯度回波读出进行成像。该序列在体模实验、有和没有全髋关节置换的健康志愿者体内以及接受标准 MRI 前列腺检查的患者体内进行了验证。

结果

所提出的序列对标准扩散加权 EPI 在存在中度离频时遇到的阴影和失真伪影具有鲁棒性。通过所提出的序列获得的表观扩散系数估计值与通过扩散加权 EPI 获得的值相当。

结论

在常规全身 3T MRI 上,使用 800 的 b 值和名义分辨率为 1.7 1.7 4mm 的条件下,在 6 分钟的扫描时间内,对全髋关节置换患者的前列腺进行无失真弥散加权成像的采集是可行的。