Comparison of the Accuracy and Reliability of Instrumental Shade Selection Devices and Visual Shade Selection: An in Vitro Study.

作者信息

Kutkut N, Jordi M, Almalki A, Conejo J, Anadioti E, Blatz M

机构信息

Department of Preventive and Restorative Dentistry, University of Pennsylvania, Philadelphia, Pennsylvania, USA.

Department of Restorative Dentistry, University of Siena and Marseille University, Siena, Italy.

出版信息

J Esthet Restor Dent. 2025 Feb;37(2):477-484. doi: 10.1111/jerd.13311. Epub 2024 Sep 2.

Abstract

OBJECTIVE

New shade-matching devices are available for clinical use, but their accuracy and reliability have not been investigated. This in vitro study compared the accuracy and reliability of a shade-matching colorimeter (Optishade), a spectrophotometer (VITA Easyshade V), and visual shade matching in simulated clinical conditions.

MATERIAL AND METHODS

Shade matching was conducted with two digital devices; (Optishade and Vita Easyshade) and the visual shade selection method, with a new 16-tab Vita Classical shade guide. Shade tabs were placed within a gingival matrix and measurements were taken from the central area of each shade tab enclosed in a black box to simulate the oral cavity. Visual assessments were conducted independently by two examiners. For the reliability assessment, each shade tab was measured 20 times by each examiner. Accuracy was calculated as a percentage of the matching readings with the actual shade per each tab. Data were analyzed with the one-way ANOVA test (α = 0.05) and Pairwise Comparison (Bonferroni Post hoc Test) to assess accuracy. The reliability evaluation was conducted by comparison between both examiners using mean and standard deviation; by a two-tailed t-test for the independent samples for all the three groups included in the study.

RESULT

The shade matching accuracy of Optishade was 97.50%, Vita Easyshade 89.38%, and visual method 62.19%. Statistically, high significant difference in accuracy (p < 0.001) was found when comparing both Optishade and VITA Easyshade with the visual method. Additionally, there was statistically significant difference in accuracy (p < 0.05) when Optishade was compared to Vita Easyshade. The reliability of the Optishade was 99.6% and Vita Easy shade was 99.38%. Statistically, showed no statistically significant difference in the reliability of shade match obtained within the same group when comparing the readings of Examiner 1 and Examiner 2 (p > 0.05).

CONCLUSION

The colorimeter demonstrated the highest accuracy, followed by the spectrophotometer, and the visual shade selection method. Both digital methods investigated in this study showed comparable reliability.

CLINICAL IMPLICATION

The results of this study emphasize the clinical advantage of digital shade-matching devices over visual shade-matching. The colorimeter offers high accuracy for shade matching and should be considered to use for shade matching especially with restorations in the esthetic zone.

摘要

目的

新型比色设备已可用于临床,但尚未对其准确性和可靠性进行研究。本体外研究在模拟临床条件下比较了比色色度计(Optishade)、分光光度计(VITA Easyshade V)和视觉比色的准确性和可靠性。

材料与方法

使用两种数字设备(Optishade和Vita Easyshade)以及视觉比色法,采用新的16片式Vita经典比色板进行比色。将比色片置于牙龈基质内,从置于黑盒中的每个比色片的中心区域进行测量,以模拟口腔。由两名检查者独立进行视觉评估。为进行可靠性评估,每位检查者对每个比色片测量20次。准确性计算为每个比色片与实际色号匹配读数的百分比。使用单因素方差分析(α = 0.05)和成对比较(Bonferroni事后检验)分析数据以评估准确性。通过比较两名检查者的均值和标准差进行可靠性评估;对研究中纳入的所有三组独立样本进行双尾t检验。

结果

Optishade的比色准确性为97.50%,Vita Easyshade为89.38%,视觉方法为62.19%。在统计学上,将Optishade和VITA Easyshade与视觉方法进行比较时,发现准确性存在高度显著差异(p < 0.001)。此外,将Optishade与Vita Easyshade进行比较时,准确性也存在统计学显著差异(p < 0.05)。Optishade的可靠性为99.6%,Vita Easyshade为99.38%。在统计学上,比较检查者1和检查者2的读数时,同一组内获得的比色可靠性无统计学显著差异(p > 0.05)。

结论

色度计显示出最高的准确性,其次是分光光度计,然后是视觉比色法。本研究中调查的两种数字方法显示出相当的可靠性。

临床意义

本研究结果强调了数字比色设备相对于视觉比色的临床优势。色度计在比色方面具有很高的准确性,尤其在美学区域进行修复时,应考虑使用其进行比色。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验