Mapleson-D系统各种改良方式下的重复呼吸评估。

Evaluation of rebreathing in various modifications of the Mapleson-D system.

作者信息

Campbell R L, Lee N R, Shamaskin R G, Priest J H, Rah K H

出版信息

J Oral Maxillofac Surg. 1985 Aug;43(8):574-9. doi: 10.1016/0278-2391(85)90123-5.

Abstract

Several modifications of the Bain system for nitrous oxide analgesia or general anesthesia were tested for rebreathing and retention of CO2. The classic Bain circuit causes the most rebreathing as measured by the fractional concentration of CO2 in a sample of inspired gas (FICO2), but stimulation of the respiratory system seems to allow near normal end tidal CO2 and arterial CO2 tensions. The most predictable systems that have the least amount of rebreathing are those that have the least dead space.

摘要

对用于氧化亚氮镇痛或全身麻醉的贝恩系统的几种改进进行了测试,以观察其对二氧化碳的再呼吸和潴留情况。经典的贝恩回路在吸入气体样本中的二氧化碳分数浓度(FICO2)测量中导致最多的再呼吸,但呼吸系统的刺激似乎能使呼气末二氧化碳和动脉二氧化碳张力接近正常。再呼吸量最少且最可预测的系统是死腔量最少的系统。