在新型滑环式放射治疗系统上采用新型后处理算法对千伏级X射线图像引导系统进行特性描述。

Characterization of kilovoltage x-ray image guidance system with a novel post-processing algorithm on a new slip ring-mounted radiotherapy system.

作者信息

Zhu Heling, Dong Tingting, Pang Tingtian, Guan Qiu, Yang Jingru, Zhao Feini, Yang Bo, Qiu Jie

机构信息

Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.

Our United Corporation, Xi'an, Shanxi Province, China.

出版信息

J Appl Clin Med Phys. 2024 Dec;25(12):e14524. doi: 10.1002/acm2.14524. Epub 2024 Sep 11.

Abstract

PURPOSE

This study evaluates the performance of a kilovoltage x-ray image-guidance system equipped with a novel post-processing optimization algorithm on the newly introduced TAICHI linear accelerator (Linac).

METHODS

A comparative study involving image quality tests and radiation dose measurements was conducted across six scanning protocols of the kV-cone beam computed tomography (CBCT) system on the TAICHI Linac. The performance assessment utilized the conventional Feldkamp-Davis-Kress (FDK) algorithm and a novel Non-Local Means denoising and adaptive scattering correction (NLM-ASC) algorithm. Image quality metrics, including spatial resolution, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR), were evaluated using a Catphan 604 phantom. Radiation doses for low-dose and standard protocols were measured using a computed tomography dose index (CTDI) phantom, with comparative measurements from the Halcyon Linac's iterative CBCT (iCBCT).

RESULTS

The NLM-ASC algorithm significantly improved image quality, achieving a 300%-1000% increase in CNR and SNR over the FDK-only images and it also showed a 100%-200% improvement over the iCBCT images from Halcyon's head protocol. The optimized low-dose protocols yielded higher image quality than the standard FDK protocols, indicating potential for reduced radiation exposure. Clinical implementation confirmed the TAICHI system's utility for precise and adaptive radiotherapy.

CONCLUSION

The kV-IGRT system on the TAICHI Linac, with its novel post-processing algorithm, demonstrated superior image quality suitable for routine clinical use, effectively reducing image noise without compromising other quality metrics.

摘要

目的

本研究评估了配备新型后处理优化算法的千伏级X射线图像引导系统在新引入的太极直线加速器(直线加速器)上的性能。

方法

在太极直线加速器上对千伏锥形束计算机断层扫描(CBCT)系统的六种扫描协议进行了一项涉及图像质量测试和辐射剂量测量的对比研究。性能评估采用了传统的费尔德坎普-戴维斯-克雷斯(FDK)算法和一种新型的非局部均值去噪和自适应散射校正(NLM-ASC)算法。使用Catphan 604体模评估图像质量指标,包括空间分辨率、对比度噪声比(CNR)和信噪比(SNR)。使用计算机断层扫描剂量指数(CTDI)体模测量低剂量和标准协议的辐射剂量,并与Halcyon直线加速器的迭代CBCT(iCBCT)进行对比测量。

结果

NLM-ASC算法显著提高了图像质量,与仅使用FDK的图像相比,CNR和SNR提高了300%-1000%,并且与Halcyon头部协议的iCBCT图像相比也提高了100%-200%。优化后的低剂量协议产生的图像质量高于标准FDK协议,表明有降低辐射暴露的潜力。临床应用证实了太极系统在精确和自适应放射治疗中的实用性。

结论

太极直线加速器上的千伏级图像引导放射治疗(IGRT)系统及其新型后处理算法展示了适用于常规临床使用的卓越图像质量,在不影响其他质量指标的情况下有效降低了图像噪声。