深部脑超声消融热剂量建模及实验验证

Deep Brain Ultrasound Ablation Thermal Dose Modeling with Experimental Validation.

作者信息

Zhao Zhanyue, Szewczyk Benjamin, Tarasek Matthew, Bales Charles, Wang Yang, Liu Ming, Jiang Yiwei, Bhushan Chitresh, Fiveland Eric, Campwala Zahabiya, Trowbridge Rachel, Johansen Phillip M, Olmsted Zachary, Ghoshal Goutam, Heffter Tamas, Gandomi Katie, Tavakkolmoghaddam Farid, Nycz Christopher, Jeannotte Erin, Mane Shweta, Nalwalk Julia, Burdette E Clif, Qian Jiang, Yeo Desmond, Pilitsis Julie, Fischer Gregory S

机构信息

Worcester Polytechnic Institute, Worcester, MA.

Department of Neurosurgery, Albany Medical Center, Albany, NY.

出版信息

ArXiv. 2024 Sep 5:arXiv:2409.02395v2.

Abstract

Intracorporeal needle-based therapeutic ultrasound (NBTU) is a minimally invasive option for intervening in malignant brain tumors, commonly used in thermal ablation procedures. This technique is suitable for both primary and metastatic cancers, utilizing a high-frequency alternating electric field (up to 10 MHz) to excite a piezoelectric transducer. The resulting rapid deformation of the transducer produces an acoustic wave that propagates through tissue, leading to localized high-temperature heating at the target tumor site and inducing rapid cell death. To optimize the design of NBTU transducers for thermal dose delivery during treatment, numerical modeling of the acoustic pressure field generated by the deforming piezoelectric transducer is frequently employed. The bioheat transfer process generated by the input pressure field is used to track the thermal propagation of the applicator over time. Magnetic resonance thermal imaging (MRTI) can be used to experimentally validate these models. Validation results using MRTI demonstrated the feasibility of this model, showing a consistent thermal propagation pattern. However, a thermal damage isodose map is more advantageous for evaluating therapeutic efficacy. To achieve a more accurate simulation based on the actual brain tissue environment, a new finite element method (FEM) simulation with enhanced damage evaluation capabilities was conducted. The results showed that the highest temperature and ablated volume differed between experimental and simulation results by 2.1884°C (3.71%) and 0.0631 cm (5.74%), respectively. The lowest Pearson correlation coefficient (PCC) for peak temperature was 0.7117, and the lowest Dice coefficient for the ablated area was 0.7021, indicating a good agreement in accuracy between simulation and experiment.

摘要

体内基于针的治疗性超声(NBTU)是干预恶性脑肿瘤的一种微创选择,常用于热消融手术。该技术适用于原发性和转移性癌症,利用高频交变电场(高达10兆赫)激发压电换能器。换能器产生的快速变形会产生一种声波,该声波在组织中传播,导致目标肿瘤部位局部高温加热并诱导细胞快速死亡。为了优化NBTU换能器在治疗期间热剂量传递的设计,经常采用对变形压电换能器产生的声压场进行数值建模。由输入压力场产生的生物热传递过程用于跟踪施药器随时间的热传播。磁共振热成像(MRTI)可用于通过实验验证这些模型。使用MRTI的验证结果证明了该模型的可行性,显示出一致的热传播模式。然而,热损伤等剂量图对于评估治疗效果更具优势。为了基于实际脑组织环境实现更准确的模拟,进行了一种具有增强损伤评估能力的新有限元方法(FEM)模拟。结果表明,实验结果与模拟结果之间的最高温度和消融体积分别相差2.1884°C(3.71%)和0.0631厘米(5.74%)。峰值温度的最低皮尔逊相关系数(PCC)为0.7117,消融区域的最低骰子系数为0.702,1,表明模拟与实验在准确性上具有良好的一致性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/dbbc/11398545/2d036f40ee7f/nihpp-2409.02395v2-f0001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验