ARTORG Center for Biomedical Engineering Research, University of Bern, Bern, Switzerland.
Ann Biomed Eng. 2024 Dec;52(12):3196-3207. doi: 10.1007/s10439-024-03584-z. Epub 2024 Sep 17.
The hemodynamic performance of different prosthetic heart valves is difficult to compare among studies due to a variety of test conditions and experimental techniques. Existing studies are typically limited to one family of valves (biological or mechanical) and testing conditions of 5l/min and often lack sufficient spatial resolution. To address these limitations, a pulse duplicator with a multi-view imaging system (Tomo-PIV) was employed to investigate the three-dimensional flow field in the aortic root of three different valves: a tri-leaflet mechanical heart valve (TRIFLO, Novostia), a bi-leaflet mechanical heart valve (On-X, Artivion), and a biological heart valve (Perimount, Edwards Lifesciences). The valves were tested at low (3 l/min), normal (5 l/min), and elevated (7 l/min) cardiac output under hypotensive (40/60mmHg), normotensive (80/120mmHg), and moderate hypertensive (105/170mmHg) pressure conditions, respectively. Compared to the Perimount, peak mean velocity was - 33%, - 24%, - 18% for the TRIFLO and - 32%, - 20%, - 11% for the On-X at low, moderate, and elevated , respectively. Corresponding peak values decreased by - 66%, - 57%, - 44% (TRIFLO) and - 60%, - 50%, - 36% (On-X). At low , was lower for Perimount (1.07cm) than for TRIFLO (1.47cm) and On-X (1.52cm), while it increased for elevated to 2.75cm (TRIFLO) and 2.16cm (Perimount and On-X). For all valves, increasing led to increased flow velocities, higher and higher levels of turbulence, and the spatial influence of the valve on the flow field in the ascending aorta was extended. peaked closer to the STJ than for TRIFLO and Perimount.
由于各种测试条件和实验技术的影响,不同人工心脏瓣膜的血流动力学性能难以在研究中进行比较。现有研究通常仅限于一种瓣膜(生物瓣或机械瓣)和 5l/min 的测试条件,并且往往缺乏足够的空间分辨率。为了解决这些限制,采用具有多视角成像系统(Tomo-PIV)的脉动复制仪来研究三种不同瓣膜在主动脉根部的三维流场:三叶机械心脏瓣膜(TRIFLO,Novostia)、双叶机械心脏瓣膜(On-X,Artivion)和生物心脏瓣膜(Perimount,Edwards Lifesciences)。在分别分别为低血压(40/60mmHg)、正常血压(80/120mmHg)和中度高血压(105/170mmHg)条件下,以低(3l/min)、正常(5l/min)和高(7l/min)心输出量对瓣膜进行测试。与 Perimount 相比,在低、中、高 时,TRIFLO 的峰值平均速度分别为-33%、-24%、-18%,而 On-X 的峰值平均速度分别为-32%、-20%、-11%。相应的峰值 值分别减少了-66%、-57%、-44%(TRIFLO)和-60%、-50%、-36%(On-X)。在低 时,Perimount(1.07cm)的 低于 TRIFLO(1.47cm)和 On-X(1.52cm),而在高 时增加到 2.75cm(TRIFLO)和 2.16cm(Perimount 和 On-X)。对于所有瓣膜,随着 的增加,流速增加, 和湍流水平增加,并且瓣膜对升主动脉流场的空间影响得到扩展。 在 STJ 附近达到峰值,而不是像 TRIFLO 和 Perimount 那样。