Houdek P V, Fayos J V, Van Buren J M, Ginsberg M S
Med Phys. 1985 Jul-Aug;12(4):469-72. doi: 10.1118/1.595673.
A stereotaxic radiotherapy technique that permits accurate delivery of highly localized dose to a small intracranial target has been developed. The technique facilitates precise integration of the diagnostic and therapeutic procedures including target localization, treatment planning, simulation, repetitive patient irradiation, and daily treatment verification. A conventional linear accelerator and computed tomography scanner as well as special diagnostic and therapeutic guides are used. A suitable dosimetric distribution is achieved using arc therapy with small radiation fields and 10-MV x rays.
一种立体定向放射治疗技术已被研发出来,该技术能够将高度局部化的剂量精确地投送至颅内小靶点。这项技术有助于诊断和治疗程序的精确整合,包括靶点定位、治疗计划制定、模拟、患者重复照射以及每日治疗验证。使用了一台传统的直线加速器和计算机断层扫描仪以及特殊的诊断和治疗导向装置。通过采用小辐射野的弧形治疗和10兆伏的X射线,实现了合适的剂量分布。